58
Error: locating logo.bmp
Ensure that the USB cable is
connected properly. Ensure that the
logo file is present on the USB drive
and it is called “logo.bmp.”
Error: upgrading software
Insert the upgrade dongle in the save
screen.
Application has encountered a critical
error and is shutting down.
Please reboot the system.
No live video on preview
• Ensure that the camera is on and
check the cable connections.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE®, ensure that SIDNE is
powered on.
No live video on the monitor
• Check the cable connections
• Ensure that the camera is powered
on.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE, ensure that SIDNE is
powered on.
No live video in PIP on preview
• Check the cable connections
• Ensure that the camera is powered
on.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE, ensure that SIDNE is
powered on.
• Ensure that you are in PIP mode.
No live video PIP on monitor
• Check the cable connections.
• Ensure that the camera is powered
on.
• Ensure that the proper video source
is selected.
• If the video is routed through
SIDNE®, ensure that SIDNE is
powered on.
• Ensure that the monitor OSD PIP is
enabled via the application.
Содержание 0240050988
Страница 1: ...SDC Ultra 0240050988...
Страница 2: ......
Страница 4: ......
Страница 42: ...38 1 1 This shows the playback getting started 1 1 This shows the playback in progress...
Страница 43: ...39 1 1 This shows the playback is paused...
Страница 53: ...49 User Management Add Remove Users 1 1 Enable security for the system DICOM Settings 1 1 DICOM connection name...
Страница 55: ...51 SCU Setting 1 1 Service Class User SCU Settings 1 1 Shows all mandatory fields as defined by the PACS server...
Страница 73: ......