3.4.3 Switch on the vacuum
a)
b)
Pressure level 1 LED will turn on
Press the I/O
button to release
vacuum
Select vacuum level.
Press the appropriate key to select a vacuum level.
The selected vacuum level will be displayed by light emitting diodes.
Pressure allocation (standard programming)
Level 1 for smooth skin
Level 2 for slightly hairy skin
Level 3 for moderately hairy skin
Level 4 for very hairy skin
Note:
The lower the vacuum level selected, the better it will be tolerated by the skin.
(Look at safety instructions)
3.4.4 Applying the vacuum electrodes
First, spray individual spots with electrode spray on skin.
Note
-
Use only Signa Spray
.
-
Spray only individual spots, in order to prevent
short circuits between electrodes.
-
The electrodes may only be applied to unwounded intact skin.
-
Conductive parts of the electrode leads and connected plug-in devices of the
application parts, including the electrode, must not touch other conductive
parts, including earth!
The electrodes are applied by touching the code plates.
This will switch on the vacuum at the electrode.
In order to ensure artefact-free measurement, take care
that there is no tension acting on the electrode lines
when applying the electrodes.
3.4.5 Switch off the vacuum
To switch off the vacuum press the I/O button.
Subsequently the blow-out feature will be started.
-
The electrodes will go off automatically.