3.4
Operation
Press power switch to switch
on the vacuum unit
(green light will turn on)
3.4.1 Operation - Starting
Push the ON/OFF button at the control unit to start operation.
All LEDs of the control unit will be lit. Simultaneously, the blow-out
mode will be triggered for approximately 5 seconds.
Air will be blown through the electrodes with increased power
in order to remove any residual moisture.
Starting at the highest pressure level, the LEDs will turn off
subsequently, until all but the one for the lowest pressure level
are off. Alternating flashing of the LED associated with the selected pressure level
and simultaneous flashing of the remaining LEDs will signal general readiness of the device.
3.4.2 Position the support arm as desired
a) move up the arm
b) move the arm downward
Note:
The support arm should be adjusted to a position in which
-
the patient will not collide with the support arm
-
the electrodes can be applied without tension
-
the vacuum lines will not be bent
open the flap, position the telescopic arm,
close the flap
b
)
Example for a lower arm position:
Simply push down the arm until the desired
position is reached.
a) Example for a higher arm position:
When moving up the arm, a clicking in
regular intervals is heard. These are the rest
positions for the vertical adjustment of the
arm (if the arm is completely pulled out, it
automatically can drop maximally to such a
position!).