118
Error signals / troubleshooting
Error
code
Cause
Remedy
204
Computation error
Repeat the last measure-
ment/measuring procedure
208
Reception signal
of laser too weak,
distance above
70 m
Use an alternative target
209
Reception signal of
laser too strong
Reflections on the target
surface. Use an alternative
target. Reduce the distance
252
Temperature too
high
Cool the device
253
Temperature too
low
Warm up the device to the
required operating tempe-
rature
255
Hardware error
Switch the device off and
back on. If the error mes-
sage is still shown, contact
our service hotline.