10.2.1
On the LCD handset
10.2.2
On the control box
10.3
Emergency adjustments: Lowering the backrest fast
WARNING
1
Hold the top end of the LCD handset
briefly (for about ½ second) against
the orange magnetic unlocking key to
gain access to the staff control level
on the LCD handset.
2
The CPR adjustment function is pre-
set.
3
Press the UP or DOWN button until the
bed has reached the desired position.
1
Press the unlock button and the CPR button at the same
time. The bed moves to the desired position at maximum
speed.
In an emergency, the backrest can be lowered by hand. The operating lever for emergency
lowering is used for this purpose. The adjustment speed for lowering the backrest in an
emergency depends on the mattress on the bed and the weight of the patient.
Risk of injury
Failure to heed this warning may result in patients and other people being exposed to dan-
ger. A heavy patient can cause the backrest to drop suddenly when the red operating lever
is used.