Operating Instructions
If supply is low or has run out, install new container and prime
associated chemical pump. Refer to
for detergent container replacement
instructions and to
,
for priming
procedure.
IMPORTANT:
DO NOT insert suction tube into container without
verifying it is for the proper application (refer to S
, for details).
11. To achieve maximum cleaning efficiency, select detergents
appropriate to soil type being processed.
NOTE: For cleaning and descaling products, note the following:
1) Always use a non-foaming detergent for effective cleaning
and proper pump and water-level operation.
2) Refer to Detergent Specifications in S
, and follow
detergent manufacturer's recommendations to determine the
temperature of the Wash treatment. Refer to
Table 5-1,
Cycle Description Chart, for available temperature ranges.
3) Follow detergent manufacturer's recommendations for
amount of detergent to use depending on water hardness.
4) Contact STERIS for information on specific cleaning and des-
caling products recommended for use with this equipment.
4.2 Priming Procedure
1.
Open chamber access door.
2.
Press
CYCLE MENU
touch pad to scroll to third Cycle menu.
3.
Press
SELECT CYCLE
touch pad to select PRIME:
__ indicates flashing position
4.
Press
CYCLE/START
touch pad to start priming. Display shows:
__ indicates flashing position
5.
Using
VALUE arrows
(up or down), scroll through pump names
until pump to be primed is displayed.
6.
Press
CYCLE/START
touch pad to confirm pump selection. Dis-
play shows:
CAUTION – POSSIBLE EQUIP-
MENT DAMAGE:
• Always position each mani-
fold and/or bottom rotary
spray over a manifold con-
nector before operating unit.
If manifolds and/or bottom
rotary sprays are not posi-
tioned correctly, damage will
result and unit will be unable
to effectively wash load.
• When choosing a detergent,
select one with a low-chlo-
ride content. Detergents with
a high-chloride content must
not be used, as such deter-
gents may harm stainless
steel.
CYCLE 9
CYCLE 10
DESCALER
PRIME
PUMP TO PRIME?
PUMP X
PRESS AND HOLD START
TO PRIME (Pump Name)
WARNING – CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
Detergents are caustic and
can cause adverse effects to
exposed tissues. Do not get in
eyes, on skin or attempt to
swallow. Read and follow the
precautions and instructions
on the detergent label and in
the Material Safety Data Sheet
(MSDS) prior to handling the
detergent, refilling the deter-
gent container or servicing
the detergent injection pump
and lines. Wear appropriate
Personal Protective Equip-
ment (PPE) whenever han-
dling the detergent or
servicing the detergent injec-
tion pump and lines.
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...