5-7
Cycle Programming
Operator Manual
122997-341
11. Press
CHANGE VALUES
touch pad. Display shows Acid Soak
screen:
CYCLE 1
ACID SOAK
T=MM:SS
__indicates flashing position.
12. To enter acid soaking time, press
CURSOR (left or right)
touch
pad to select position and
VALUE (up or down)
touch pad to
select desired number (0-9). Injection time is input as minutes
and seconds within a range of 0-99 minutes and 0-99 seconds.
13. Press
CHANGE VALUES
touch pad. If option applies, display
shows Acid Wash Water Reuse option:
SAVE ACID WASH WATER
YES
__indicates flashing position.
NOTE: Question is displayed only if Acid tank option is present
on unit.
14. Press
VALUE (up or down)
touch pad to toggle between YES
and NO answer. Selecting NO deactivates Acid wash water
reuse option. Selecting YES activates Acid wash water reuse
option, so that at end of treatment, water is sent back to Acid
tank (option).
5.3.4 Rinse
1.
Press
CHANGE VALUES
touch pad. Rinse 1 phase values
screen appears:
CYCLE 10
SP=180.0F
Rinse 1
T= 01: 30
__indicates flashing position.
2.
To enter First Rinse temperature setpoint and phase time, press
CURSOR (left or right)
touch pad to select position and
VALUE
(up or down)
touch pad to select desired number (0-9). Temper-
ature setpoint is input as any number value to 1/10 of a degree.
Phase time is input as minutes and seconds within a range of 0-
49 minutes and 0-99 seconds. Temperature is adjustable from
70 to 185 °F (22 to 85°C).
3.
Press
CHANGE VALUES
touch pad. Rinse 1 Temperature Guar-
antee option screen appears:
GUAR. RINSE 1 TEMP.
NO
__indicates flashing position.
4.
Press
VALUE (up or down)
touch pad to toggle between YES or
NO answer. Selecting NO starts First Rinse phase time count-
down at beginning of phase. Selecting YES guarantees that First
Rinse phase time counts down only when water temperature is
equal to or greater than programmed setpoint.
CHANGE
VALUES
CURSOR
VALUE
CHANGE
VALUES
VALUE
CHANGE
VALUES
CURSOR
VALUE
CHANGE
VALUES
VALUE