5-5
Cycle Programming
Operator Manual
122997-341
5.
Press
CHANGE VALUES
touch pad. If option applies, display
shows Conductivity Option screen:
SELECT CONCENTRATION
1/4 OZ
__indicates flashing position.
6.
Press
VALUE
(up or down) touch pad to scroll available concen-
trations (1/4, 1/2, 1, 2 oz and default concentration name).
7.
Press
CHANGE VALUES
touch pad. If option applies, display
shows Time-based Alkaline Neutralization Option screen:
CYCLE 10
ACID INJ.
T=MM:SS
__indicates flashing position.
NOTE: If water is saved, Return to Alkaline Tank question is not
displayed. To enter alkaline neutralizer injection time, press
CURSOR touch pad to select position and VALUE touch pad to
select desired number (0-9). Injection time is input as minutes
and seconds within a range of 0-99 minutes and 0-99 seconds.
8.
Press
CHANGE VALUES
touch pad. Display shows Alkaline
wash water reuse option:
SAVE ALK. WASH WATER
YES
__indicates flashing position.
9.
Press
VALUE
(up or down) touch pad to toggle between YES
and NO answer. Selecting NO deactivates Alkaline wash water
reuse option. Selecting YES activates Alkaline wash water reuse
option, so that at end of treatment, water is sent back to Alkaline
tank.
5.3.3 Acid Wash
1.
Press
CHANGE VALUES
touch pad. Acid wash phase values
screen appears:
CYCLE 10
SP=140.0F
ACID WASH
T=MM:SS
__indicates flashing position.
NOTE: If treatment time is set to 0, treatment is bypassed and
question is not displayed.
2.
To enter Wash temperature setpoint and phase time, press
CUR-
SOR (left or right)
touch pad to select position and
VALUE (up
or down)
touch pad to select desired number (0-9). Tempera-
ture setpoint is input as any number value to 1/10 of a degree,
between 70 and 185°F (22 and 85°C). Phase time is input as
minutes and seconds within a range of 0-49 minutes and 0-99
seconds.
CHANGE
VALUES
VALUE
CHANGE
VALUES
CHANGE
VALUES
VALUE
CHANGE
VALUES
CURSOR
VALUE