Installation Checklist
Operator Manual
122997-341
❑
Washer is positioned, as shown on equipment drawing, with
required service clearance space and in relation to building sup-
ply lines.
❑
Basil 4600 Cage and Rack Washer
and
Basil 4602 Cage and
Rack Washer
must be installed between two walls, with a key-
locked service door, so washer service side is not accessible to
operator.
❑
Building steam line provides maximum dynamic steam pressure
and flow rate to washer as specified on equipment drawing.
❑
Drip leg with steam trap installed in steam supply line.
❑
Building hot water line supplies water to washer at pressure and
temperature specified on equipment drawing.
❑
Building cold water line supplies water to washer at pressure
specified on equipment drawing.
❑
Electrical supply for washer is as specified on equipment draw-
ing.
❑
Condensate returns are sized as specified on equipment draw-
ing.
❑
Vent connections are sized as specified on equipment drawing.
❑
Recirculation pump pressure is within 25 to 60 psig.
❑
Recirculation pump motor rotating in direction shown by arrow.
❑
Self-cleaning screen assembly functioning properly.
❑
Carriage drive motor rotating in direction shown by arrow.
❑
Carriage drive system functioning properly.
❑
Carriage drive motor amperage within rating indicated on the
motor.
❑
Optional exhaust fan rotating in direction shown by arrow.
❑
All piping is leak-free.
❑
Chamber sump steam coil functioning properly.
❑
Door safety switch(es) functioning properly.
❑
Cabinet joints are completely sealed, no leaks (for verification,
run machine for 1/2 hour).
❑
Door(s) easily opens from inside of chamber.
❑
Safety cables immediately stop washer operation when pulled.
❑
Each gear box plastic cap removed and replaced with air vent
provided.
❑
Floor surrounding unit has non-slip surface.
IMPORTANT:
After a few weeks of operation, inspect unit for leaks.
Retighten all clamps and connections.