2-2
122998-741
Operator Manual
Installation
2.1.2 Permissible
Environmental Conditions
This equipment is designated to give optimal results under the fol-
lowing conditions:
Indoor use only;
Altitude of operation up to: 6,267 ft (2,000 m);
Maximum relative humidity is 80% for temperatures up to: 88ºF
(31ºC) decreasing linearly to 50% relative humidity at 104ºF
(40ºC);
Pollution degree 2.
2.1.3 Seismic Anchorage
System
A Seismic Anchorage System is available for high risk seismic
zones.
2.2 Installation
Checklist
After your washer has been installed by qualified service techni-
cians, complete the following checklist to assure installation is com-
plete and correct. If you desire, contact STERIS and schedule a
technician to test installation and demonstrate proper equipment
operation.
❑
To allow service of unit without shutting off building supply lines,
shutoff valves (not provided by STERIS) for maintenance pur-
poses, are installed on steam and water lines to unit. Shutoff
valves must be capable of being locked in OFF position only.
❑
Disconnect switches (not provided by STERIS) are installed in
electric supply lines near washer and within 10' (3 m) of electri-
cal control box. Disconnect switches must be capable of being
locked in OFF position only.
NOTE: If washer is installed next to other equipment, shut-off
valves and disconnect switch should be placed so that service
can be shut off to one piece of equipment at a time.
❑
Washer is positioned as shown on equipment drawing with
required service clearance space and in relation to building sup-
ply lines.
❑
Washer must be installed between two walls with a key-locked
service door so service side is not accessible to operators.
❑
Building steam line provides maximum dynamic steam pressure
and flow rate to washer as specified on equipment drawing.
❑
Drip leg with steam trap installed in steam supply line.
❑
Building hot water line supplies water to washer at pressure and
temperature specified on equipment drawing.
❑
If applicable, building cold water line supplies water to washer at
pressure specified on equipment drawing.
❑
Electrical supply for washer is as specified on equipment draw-
ing.