Fault

Possible cause

Action

The mixer does not start

Fault in power supply

Check power supply

The impeller does not rotate

Male bearing or magnetic drive not
mounted

Dismount the drive unit, detach the
impeller and mount the male bear-
ing, impeller and magnetic drive

Insufficient power

Motor connected incorrectly

Check wiring diagram and wiring

Poor mixing

a) The impeller does not rotate at
the maximum speed
b) Impeller rotating in the wrong di-
rection

a) Increase speed, pay attention to
the guidelines in section 3.6.2.
b) Check the drive installation

AC: phase reverse power cables

DC: pole reverse power cables

AIR: swap the air supply and outlet

hoses.

Verify the impeller rotates clockwise
when seen from above

The mixer does not reach the re-
quired maximum rpm

Incorrectly set parameters in the

VFD (Variable Frequency Drive)

Adjust and calibrate the VFD

Slow and inefficient compressed air
driven motor

Clogged motor housing

Flush with paraffin as per motor
supplier’s maintenance instructions

Noise from the mixer

a) Impeller has rotated without fluid
between the male and female bear-
ing
b) Worn out bearings
c) Missing male bearing
d) Mechanical interference (con-
tact) between impeller and weld
plate
e) Mechanical interference (con-
tact) between magnet rotor and

weld plate

a) Stop the mixer. Do not restart un-
til the fluid level covers the impeller
b) Replace the bearing and the im-
peller
c) see b)
d) Reinstall the weld plate correctly
e) Realign the magnet rotor

Noise from the drive unit (first dis-
card that this is not ’normal noise’
amplified by the vessel acting as a
soundboard)

a) New drive unit has not been al-
lowed to run in b) Gearbox prob-
lems. Running insufficiently lubri-
cated

a) Run the drive unit for a couple
of hours to see if the noise disap-
pears b) Refer to gearbox supplier
maintenance manual

Magnetic coupling disconnected

a) Acceleration too rapid
b) Speed to high for the application

a) Adjust acceleration
b) lower speed.

Insufficient cleaning of the impeller

Poor fluid flow through the impeller
a) Due to low fluid level
b) Due to low speed
c) Due to high speed
d) Magnetic particles

a) Increase fluid level
b) Increase spead
c) Reduce speed and the formation
of vortices
d) Magnetic particles are not re-
moved during normal cleaning. The
impeller must be removed and
cleaned separately

n

Table 7

Troubleshooting.

14

|

© Steridose

www.steridose.com

Содержание Sanimixer

Страница 1: ...4 Mounting and dismounting the drive unit 5 3 5 Connection to main power 6 3 6 Installation check and calibration 6 4 Operation 7 4 1 Allowed operating temperature range 7 4 2 Allowed pressure range 7 4 3 Chemical compatibility 7 4 4 Operational limits 7 4 5 Further operational advice 9 4 6 Guidance for sterilizing Sterim ixer Sanimixer 10 4 7 Specific information for the use of Sterim ixer Sanimi...

Страница 2: ... or indirectly create risks to the safety and health of the person undertaking the work or the Sterim ixer Sanimixer and or its components while the work is being undertaken It is the responsibility of the user to ensure that appro priate controls and precautions are identified and applied in relation to the work covered by this document in ac cordance with relevant statutory legal and industry re...

Страница 3: ...er is not to be in stalled immediately the Sterimixer Sanimixer should be repacked in the original package and stored in a suit n Figure 1 Mounting the bearing unit on the weld plate in the vessel The figure shows the bearing unit and the PTFE washer Use a torque wrench refer to table 2 for recommended torque able storage after receipt and inspection The following should be considered with regards...

Страница 4: ...ve unit and the magnet rotor are installed Failure to do so will cause a sudden attraction between the mounted rotor and the impeller causing the latter to violently hit the bearing assembly with likely bearing damage to occur CAUTION Before mounting the impeller make sure that the drive unit is NOT mounted on the vessel NOTICE To facilitate mounting and dismounting the impeller Steridose has deve...

Страница 5: ... driven gearbox flange and rotor some differences occur depending on type of drive and size Be certain that you can handle the weight of the drive unit Take a firm grip of the gearbox to balance the weight of the drive unit and to minimize the risk of damage caused by crushing when dismounting Use appropri ate lifting device for the larger models SM 120 190 and larger 3 4 1 Mounting the drive unit...

Страница 6: ...For AC drive applications The Sterim ixer Sanimixer starts at low speed 5 Visually check the direction of rotation for the im peller which should be clockwise when seen from above If the Sterimixer Sanimixer rotates counter clockwise refer to section 6 6 Stop the Sterimixer Sanimixer immediately after completed verification CAUTION Dry operation of the Sterimixer Sanimixer will damage the bearings...

Страница 7: ...Allowed pressure range Weld plate size Design pressure range All sizes 1 7 bar g 14 5 101 5 psig 4 3 Chemical compatibility The impeller and bearing are not to be exposed to strong chemicals Nominal pH range for tungsten carbide TC n Figure 6 Sterimixer Sanimixer should never be allowed to run dry n Figure 7 The bearing lubrication principle of the SMA type impeller bearings is 2 14 but silicon ca...

Страница 8: ...s IEC ATEX NEMA XP NEMA Stainless Steel etc this may vary Bearing materiala Tungsten Carbide Silicion Carbide Dri amond Chemical compatibility pH range 2 14b Best inert Same as silicon carbide Ease of installation and handling Best Harder than tungsten car bide but more brittle Somewhat better than sili con carbide Survivability under dry running conditions Good Poor Best lowest coefficient of fri...

Страница 9: ...optimum performance in cleaning in place CIP without any need for dismounting n Figure 9 Succesful cleaning of the impeller relies on a combination of the self cleaning flow through design of the SMA type impeller the cleaning agent used and respecting the prescribed time and temperature The flow channels in the impeller provide a process fluid flow through the bearings In order to achieve this th...

Страница 10: ...e drive unit and gearbox shall not be exposed to ambient temperatures above 40 C NOTICE Use caution when exposing to temperatures over 135 C as this may decrease the life expectancy of the elastomeres used in the bearing assembly A certain delay in the rise of temperature between the inside of the impeller and the vessel is caused by the mass of the impeller head This delay is influenced by Sterim...

Страница 11: ...nverter s set value speed exceeds 10 the drive is to be disabled interlocking In this case the magnetic field between the inner and outer magnetic rotor is broken indicating a malfunction 4 7 4 Assessment Components outside of the tank are in Ex zone 1 or 2 and are delivered with the corresponding ATEX declaration of conformity from the manufacturer Ex II 2G b c de IIC T4 The machine mixer tank wi...

Страница 12: ... lubri cation creating excessive wear and heat and subse quently premature failure It is therefore important that the operator is properly trained and familiar with all procedures If the Sterimixer Sanimixer is operated beyond its operating limits severe damage might occur subse quently creating a failure It is essential that the oper ator has received specific training together with the full unde...

Страница 13: ...e a year Note that the interval will be highly dependant on the application Remove and check the bearing and the impeller for signs of surface damage and general wear of the bear ings Change the bearing and the impeller if the surface is coarse or uneven scratched or if fragments have been knocked out from the bearing The impeller can be sent to Steridose for refurbishment Change the PTFE washer w...

Страница 14: ... female bear ing b Worn out bearings c Missing male bearing d Mechanical interference con tact between impeller and weld plate e Mechanical interference con tact between magnet rotor and weld plate a Stop the mixer Do not restart un til the fluid level covers the impeller b Replace the bearing and the im peller c see b d Reinstall the weld plate correctly e Realign the magnet rotor Noise from the ...

Страница 15: ...www steridose com Installation Operation 15 ...

Страница 16: ...sk for process interference Steridose partners with the best distributors and repre sentatives in the industry all over the world Together we become the perfect mix a premium product with global references combined with local presence for product and application support Steridose AB Himmelsbodavägen 7 SE 147 39 Tumba Sweden Phone 46 8 449 9900 info steridose com www steridose com Steridose Inc 502...

Отзывы: