2.2 Installing  Steps

 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

2.2.1 

Installing the Objective (Fig. 

1-2)

 

 

1.  Adjust  the  coarse  &fine  focus  knob,  till  the 

mechanical  stage to the low limited  place. 

2. Screw  down  the objective  to nosepiece  from  left  or 

right    side,    low    magnification    objective    first. 

Install  all  the  objective  form  low  to  high 

magnification  following  the clock hand. 

  According  to  this  way  to  install  the  objective  will 

make it easier to change magnification in operation. 

 

 

Clean the objective  frequently.

 

  

At  first,  use  the  10X  objective  to  looking  for 

image,  then change  another  one.

 

  

Turn   the   objective   till   hear   the   “Kai.   Kai” 
sound,  make  sure  the  objective  enter  the 
objective  center.

 

 

 

2.2.2        Installing         the        Metallurgical

 

Illumination  Set (Fig. 3-4)

 

 

 

1.  

Insert  the  Metallurgical  Illumination  Set  with 

Lamp   House   installed    into   the Microscope 

Body.    Align  the  position  and  screw  down  the 

bolt with hexagon  wrench. 

2.  

Plug the power cord 

① 

of the Metallurgical 

Illumination  Set  into  the  socket 

.  Then 

screw the bolt firmly  on the socket. 

 

Fig. 1

 

Fig. 2

 

Fig. 3

 

Fig. 4

 

Содержание NYMCS-603

Страница 1: ...r Manual This manual is written for Metallurgical Microscope of NYMCS 603 For safety and for exerting the best performance making you familiar with the instrument entirely it is strongly recommended t...

Страница 2: ...Notices 2 1 Components Name 3 2 Installation 4 2 1 Installing Diagram 4 2 2 Installing steps 5 3 Adjustment Operation 9 3 1 Adjustment System Installing Illustration 9 3 2 Operation 11 4 Technical Sp...

Страница 3: ...pply voltage or it will bring a serious damage to the instrument 9 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to dis...

Страница 4: ...3 STEINDORFF 1 Name of Components NYMCS 603...

Страница 5: ...ing order of the spare parts The number expresses the installing process Before the installing make sure there is no dust and stain on the spare parts No marks by outside force on the surface of the s...

Страница 6: ...er to change magnification in operation Clean the objective frequently At first use the 10X objective to looking for image then change another one Turn the objective till hear the Kai Kai sound make s...

Страница 7: ...ert the binocular viewing head in the head of the body turn to the right place then fix up it by bolts 2 2 4 Install the eyepiece Fig 7 8 Insert the eyepiece in the eyepiece tube till end Fig 8 pictur...

Страница 8: ...he socket on microscope safely Make sure be connected 3 Plug the power cord into the power source Socket safely Make sure be connected Do use the supplied power cord all the time If lost or damaged se...

Страница 9: ...house and nearby parts will be very hot and will cause serious burns Please turn the main switch on O off pull up power plug and make sure the bulb the lamp house and periphery are all cool Then you...

Страница 10: ...9 STEINDORFF 3 Adjustment Operation NYMCS 603 3 1 Adjustment Set Diagram Fig 18 19 Fig 18...

Страница 11: ...10 STEINDORFF Fig 19...

Страница 12: ...just the View field Diaphragm and Aperture Diaphragm until the two Diaphragm open size is properly Then you can get clear image When in using you need to adjust simultaneously part observe until to th...

Страница 13: ...he interpupillar distance until the left and right fields of view coincide completely 3 3 6 Focusing Fig 25 1 Use the 10 objective focus to avoid the objective touch with the specimen you should raise...

Страница 14: ...Adjusting the Tension Adjustment Collar Fig 27 The tightness of the tension adjustment collar has adjusted before leaving factory if finding it s loosing the mechanical stage drop itself because of de...

Страница 15: ...Quadruple Nosepiece Objective Infinite plan Achromatic 4 10 20 choose and buy 40 Focus System Coaxial Coarse and Fine Focusing System Sensitivity and Graduation of Fine Focus 0 002mm Stage Double laye...

Страница 16: ...focus low resolution The objective damage Mend and correct the objective send to factory for overhauling The lens of the objective and eyepiece is moldy or have contaminant Do cleaning The opening of...

Страница 17: ...connect them exactly the installation of the bulb is wrong Install the bulb correctly The bulb burn out Change a new bulb 2 The bulb burn out in a high frequency Not use the specified lamp Use the re...

Отзывы: