13 

STEINDORFF

®

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Fig. 26

 

Fig. 27

 

3.3.7  

Placing Specimen (Fig. 26) 

1.

 

Place  the  slide  on  the  mechanical stage.  Use  the  stage 

clips to clamp  the slide gently.

 

2.

 

Turn the portrait and lateral adjustment knob of the 

mechanical ruler,  move  the  specimen  onto  the  required 

position.

 

 

   

Be careful  when  changing  the objective.  If you finish 

the  observation  with  the  short  working  distance 

objective, and want to change another   one,   be careful 

of not letting the objective touch the specimen.

 

 

3.3.8  

Adjusting the Tension Adjustment Collar 

(Fig. 27) 



  The   tightness   of   the   tension   adjustment   collar   has 

adjusted  before  leaving  factory,  if  finding  it’s  loosing 

(the mechanical stage drop itself because of deadweight),   

please   turning   the   tension   adjustment collar 

① 

until 

the tightness is in order. Turn it along the direction show 

in  the  picture,  the  Coarse  Focus  Knob 

② 

will become 

tighter. Turn anti-direction  will become loosen. 

If  the  mechanical stage  drop  itself,  or  even  lose  focus 

just  after  adjusting  the  fine  focus  knob 

.  Mean  the 

tightness  of the  Coarse  Focus  Knob 

② 

is  too  low.  You 

should  turn  the  tension  adjustment  collar  along  the 

direction  show  in the picture  to make  it tighter. 

Содержание NYMCS-603

Страница 1: ...r Manual This manual is written for Metallurgical Microscope of NYMCS 603 For safety and for exerting the best performance making you familiar with the instrument entirely it is strongly recommended t...

Страница 2: ...Notices 2 1 Components Name 3 2 Installation 4 2 1 Installing Diagram 4 2 2 Installing steps 5 3 Adjustment Operation 9 3 1 Adjustment System Installing Illustration 9 3 2 Operation 11 4 Technical Sp...

Страница 3: ...pply voltage or it will bring a serious damage to the instrument 9 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to dis...

Страница 4: ...3 STEINDORFF 1 Name of Components NYMCS 603...

Страница 5: ...ing order of the spare parts The number expresses the installing process Before the installing make sure there is no dust and stain on the spare parts No marks by outside force on the surface of the s...

Страница 6: ...er to change magnification in operation Clean the objective frequently At first use the 10X objective to looking for image then change another one Turn the objective till hear the Kai Kai sound make s...

Страница 7: ...ert the binocular viewing head in the head of the body turn to the right place then fix up it by bolts 2 2 4 Install the eyepiece Fig 7 8 Insert the eyepiece in the eyepiece tube till end Fig 8 pictur...

Страница 8: ...he socket on microscope safely Make sure be connected 3 Plug the power cord into the power source Socket safely Make sure be connected Do use the supplied power cord all the time If lost or damaged se...

Страница 9: ...house and nearby parts will be very hot and will cause serious burns Please turn the main switch on O off pull up power plug and make sure the bulb the lamp house and periphery are all cool Then you...

Страница 10: ...9 STEINDORFF 3 Adjustment Operation NYMCS 603 3 1 Adjustment Set Diagram Fig 18 19 Fig 18...

Страница 11: ...10 STEINDORFF Fig 19...

Страница 12: ...just the View field Diaphragm and Aperture Diaphragm until the two Diaphragm open size is properly Then you can get clear image When in using you need to adjust simultaneously part observe until to th...

Страница 13: ...he interpupillar distance until the left and right fields of view coincide completely 3 3 6 Focusing Fig 25 1 Use the 10 objective focus to avoid the objective touch with the specimen you should raise...

Страница 14: ...Adjusting the Tension Adjustment Collar Fig 27 The tightness of the tension adjustment collar has adjusted before leaving factory if finding it s loosing the mechanical stage drop itself because of de...

Страница 15: ...Quadruple Nosepiece Objective Infinite plan Achromatic 4 10 20 choose and buy 40 Focus System Coaxial Coarse and Fine Focusing System Sensitivity and Graduation of Fine Focus 0 002mm Stage Double laye...

Страница 16: ...focus low resolution The objective damage Mend and correct the objective send to factory for overhauling The lens of the objective and eyepiece is moldy or have contaminant Do cleaning The opening of...

Страница 17: ...connect them exactly the installation of the bulb is wrong Install the bulb correctly The bulb burn out Change a new bulb 2 The bulb burn out in a high frequency Not use the specified lamp Use the re...

Отзывы: