STEINDORFF

®

 

 

1) 

Optical part

 

PROBLEM 

CAUSE 

SOLUTION 

The  edge  of  the 

field of view has 

shadow or the 

brightness  is 

asymmetry 

The nosepiece  is not in the located 

position  (objective  is not in the center 

of the optical path) 

Turn objective  to the right 

position(the  optical  light 

center) 

The image of the filament  is not in the 

center 

Center  it 

The surface  of the lens is moldy or 

has contaminant  (include  condenser, 

objective,  eyepiece  and collector) 

Clean  it up 

Find  dust  and  stain  in 

the field of view

 

There are stains on the lens (include 

condenser,  objective,  eyepiece  and 

collector) 

Clean  it up 

There are stains on the specimen 

Clean  it up 

The position  of the condenser  is too 

low 

Raise  it to proper position 

The image is 

defocused (low 

resolution  \ contrast)

 

There  is no coverslip  on the 

specimen 

Add coverslip 

The coverslip  is too thick or too thin 

Use the standard  coverslip 

(0.17mm  ) 

The specimen  is placed  inversely 

Reverse  it back 

There  was oil on the dry objective 

(easily  happened  in 40X objective) 

Clean  it up 

There are stains on the lens 

(including  condenser,  objective, 

eyepiece  and collector) 

Clean  it up 

The oil objective  didn

’t immerse  oil 

Use immerse  oil 

There  was bleb in the oil 

Eliminate  the bleb 

Use a unsuitable  oil 

Change  to the specified  one 

The size of the aperture  diaphragm  is 

too large 

Minify  it 

The size of the aperture  diaphragm  is 

too small 

Open  it up 

The position  of the condenser  is too 

low 

Adjust  the position 

Содержание NYMCS-212

Страница 1: ...lar microscope NYMCS 212 Toensure the safety obtain optimum performance and to familiarize the use of this microscope it is recommended strongly that you study this manual thoroughly before using the...

Страница 2: ...NYMCS 212 is widely used in biological pathological bacteriological fields for purposes such as observation research academic demonstration clinic medical inspection It features more comfortable and...

Страница 3: ...198 3 Eyepiece Category Magnification Focal Length f mm Field of view mm Plan Eyepiece 10X 24 95 18 4 Total Magnification Eyepiece 10X 10X 10X 10X Objective 4X 10X 40X 100X Total 40X 100X 400X 1000X 5...

Страница 4: ...nosepiece has a precise frame which can locate objectives accurately and stably The digital binocular head can turn 360 degree freely with comforts and conveniences b Curved arm and coarse and fine f...

Страница 5: ...the nosepiece clockwise and mount other objectives in the sequence of low to high magnification E Condenser The condenser had been installed in the microscope before leaving factory F Collector The c...

Страница 6: ...operation In this way the specimen can come into focus rapidly just by turning the coarse focusing knob Note specimens of the same type means they are same in the thickness of slide cover slip and spe...

Страница 7: ...specimen Clean it up The position of the condenser is too low Raise it to proper position The image is defocused low resolution contrast There is no coverslip on the specimen Add coverslip The coversl...

Страница 8: ...cked position 2 Mechanical part PROBLEM CAUSE SOLUTION The image can focus when using high magnification objective The specimen is placed inversely The coverslip is too thick Turn inversely Use the st...

Страница 9: ...e digital Binocular head with PC by the USB cable the system will find the new device and then run the driver under the CD please follow the clue to finish the installation Attention please link the c...

Страница 10: ...ent USB2 0 port requires reinstalling or updating the driver again to make the camera work rightly If not it will influence the performance of the camera and cause the live image displays drag for the...

Страница 11: ...esigned for digital microscope It allows you to view capture edit record zoom measure and process microscope images Specific features included are as follows 1 Active image preview freezing preview fi...

Страница 12: ...n image zoom in zoom out 1 1 Attention When launching the software there will be several seconds displays as black preview please don t worry it is normal because the camera needs time to do some pret...

Отзывы: