5. Assembly 

A. 

Environmental  Requirement:

 

1.  Ambient  temperature:  5°C to 40°C (41°F  to 104°F), Maximum  R.H

85%. 

2.  High Temperature  will result in a mildewing,  dew and even ruinous  instrument. 

3.  Avoid placing the instrument in a dusty environment. When ending  your microscope 

operation,  please  cover it with the dust cover. 

4.  Lay the microscope  in a plan and stable position,  please. 

B. 

Input voltage  check:

 

Check the input voltage: be sure the input voltage which signed in the back of the microscope  is 

consistent  with  the power  supply  voltage,  or it will bring  a serious  damage  to the instrument. 

C. 

Eyepiece:

 

Insert the eyepiece  into the eyepiece  tube. 

D. 

Objective:

 

Adjusting the  coarse focus knob until the  mechanical stage reaches its  lower limit position. Screw 

the  lowest  magnification objective  into  the  nosepiece  from  the  left  or  the  right side, then revolve 

the nosepiece  clockwise  and mount  other objectives  in the sequence  of low to high magnification 

E. 

Condenser:

 

The condenser  had been installed  in the microscope  before  leaving  factory. 

F. 

Collector:

 

The collector  had been  installed  and located  before  leaving  factory.  The screw-mounted collector 

can be taken down by anticlockwise  turning. 

G. 

Filter:

 

Mount  the filter on the iris diaphragm  bracket  under the condenser  if necessary. 

 

6. Operation 

1.  Insert the eyepiece into the eyepiece tube and revolve the objectives into the nosepiece  in sequence 

of magnification. 

2.  Place  the specimen  in the center  of the stage, keeping  the cover  slip upward,  and then nip it with 

the specimen  holder. 

3.  Search the image by lower magnification  objective,  turn the longitudinal  and transversal adjustment 

knobs  which on the mechanical  ruler to move the specimen  onto the center of the  field  of view. 

Then  replace  with  higher  magnification  objective,  turn  the  fine  focusing knob to sharpen  the image. 

4.  When using 100X oil objective,  add the cedar oil between  the objective  and specimen. 

Содержание NYMCS-212

Страница 1: ...lar microscope NYMCS 212 Toensure the safety obtain optimum performance and to familiarize the use of this microscope it is recommended strongly that you study this manual thoroughly before using the...

Страница 2: ...NYMCS 212 is widely used in biological pathological bacteriological fields for purposes such as observation research academic demonstration clinic medical inspection It features more comfortable and...

Страница 3: ...198 3 Eyepiece Category Magnification Focal Length f mm Field of view mm Plan Eyepiece 10X 24 95 18 4 Total Magnification Eyepiece 10X 10X 10X 10X Objective 4X 10X 40X 100X Total 40X 100X 400X 1000X 5...

Страница 4: ...nosepiece has a precise frame which can locate objectives accurately and stably The digital binocular head can turn 360 degree freely with comforts and conveniences b Curved arm and coarse and fine f...

Страница 5: ...the nosepiece clockwise and mount other objectives in the sequence of low to high magnification E Condenser The condenser had been installed in the microscope before leaving factory F Collector The c...

Страница 6: ...operation In this way the specimen can come into focus rapidly just by turning the coarse focusing knob Note specimens of the same type means they are same in the thickness of slide cover slip and spe...

Страница 7: ...specimen Clean it up The position of the condenser is too low Raise it to proper position The image is defocused low resolution contrast There is no coverslip on the specimen Add coverslip The coversl...

Страница 8: ...cked position 2 Mechanical part PROBLEM CAUSE SOLUTION The image can focus when using high magnification objective The specimen is placed inversely The coverslip is too thick Turn inversely Use the st...

Страница 9: ...e digital Binocular head with PC by the USB cable the system will find the new device and then run the driver under the CD please follow the clue to finish the installation Attention please link the c...

Страница 10: ...ent USB2 0 port requires reinstalling or updating the driver again to make the camera work rightly If not it will influence the performance of the camera and cause the live image displays drag for the...

Страница 11: ...esigned for digital microscope It allows you to view capture edit record zoom measure and process microscope images Specific features included are as follows 1 Active image preview freezing preview fi...

Страница 12: ...n image zoom in zoom out 1 1 Attention When launching the software there will be several seconds displays as black preview please don t worry it is normal because the camera needs time to do some pret...

Отзывы: