3.3.10 How to use the reticle to measure?
Example
Selected the standard graduation of 0.05mm under a given power and move the micro-ruler when clearly focused
so that the coordinate corresponds with the micro-ruler. The standard ruler of every 12 graduations correspond
to 19 graduations of the coordinate ruler, then each graduations of the coordinate ruler is about 0.03mm.In this
case, user can carry out the measurement for the object observed under microscope according the distance of
coordinate ruler(requiring re-calibration of the separation of coordinate ruler each time for a changed power).
4. Outfit
4.1 Specification
Optical system
Infinite optical system
NYMCS-
1701
Reflected light
source
Excitation units
excitation
Dichroic
mirror
Barrier filter
Blue Excitation
BP460
~
490
DM500
BA520
●
Green Excitation
BP510
~
550
DM570
BA590
●
Ultraviolet Excitation
BP330
~
385
DM400
BA420
○
Violet Excitation
BP400
~
410
DM455
BA455
○
Viewing Tube
Trinocular head ,30°incline
;
interpupilary range 48-75mm
●
Eyepiece
High point ,extra-wide field eyepieceEW10×/22
●
Centering
Centering (φ30mm)
○
Nosepiece
Backward Quintuple Nosepiece
●
Objectives
In finite plan long working distance objective 4×
●
In finite plan long working distance objective 10×
○
In finite plan long working distance objective 20×
○
In finite plan long working distance objective 40×
●
In finite plan phase contrast objective PH10×
●
In finite plan phase contrast objective PH20×
●
In finite plan phase contrast objective PH40×
○
Fig.25