Error Codes and Fault Logs
14.0
30
Fault:
Reason for Fault:
Required Action:
Field Test Failed
Field test failed (Calibration).
A fault occurred during
Field Testing (calibration).
Test Flag Check
One or More Test flags Not
Set.
Redo test(s)
and/or Replace PCB.
ADC
ADC Conversion error
- Check R-net Cable,
Replace PCB Interface.
R-net UART Overflow
UART send queue is full.
Replace PCB.
R-net UART Underflow
UART receive queue is empty.
R-net Communication
Timeout
Maximum number of packet
re-transmissions is reached.
R-net Tx Overflow
Packet Transmit Buffer is Full.
R-net Rx Overflow
Packet Receive Buffer is Full
R-net Invalid Seq Nr
Received a Packet with
an Invalid Sequence Number
R-net Data Packet Error
Data Packet ACK mismatch
R-net Descr Error
Invalid Packet Data
Descriptor
R-net API Version Error
The R-net Chip Contains
an Invalid API Version
Load the latest API version
into the R-net chipset
with the R-net Dongle.
R-net Chip Tripped Error
The R-net Chip has
encountered
an Internal Error.
Internal chipset error: Replace PCB
R-net system error:
solve R-net system error.
Communication
Communication with the
sensor (joystick failed)
- Check Sensor (joystick) cable.
- Replace sensor (joystick).
Joystick (Sensor)
Faulty sensor (joystick)
Accelerometer
Accelerometer failed
Replace PCB Interface.