DATA Outputs
Data I/O Connector
The indicator has an 8 pin data I/O connector with several supported communication protocols.
These communication protocols can be accessed using a dedicated cable or through custom wiring
using the Pigtail cable.
Serial ASCII Command Protocol (USB/RS232/UART)
The gage supports several "serial" interfaces that interact with the device using a human readable
ASCII command protocol. This protocol can be accessed over USB as a virtual com (CDC) device,
RS232 or TTL level UART. For all hardware implementations, the specifications and command set
are identical.
Baud
9600
Data Bits
8
Parity
None
Stop Bits
1
Flow Control
None
Hardware Specifications
Commands
All Commands are terminated
with a new line character
<CR>. All commands and the
most common settings are
listed below. Please see
technical manual for more
details.
Examples: Enter p013=2
followed by a <CR> to change
the display units to MM.
COMMAND
DESCRIPTION
R
Print current gage reading
C
Clear gage
H
Toggle Hold Mode
rstERR
Resets errors
?pXXX
Get the value of configuration parameter "XXX"
pXXX=<new value>
Set the value of parameter "XXX" to <new value>
RESPONSES
DESCRIPTION
+ 2.1380, MM
13 character response to print reading
SUCCESS
Parameter is successfully written to
NOT FOUND
Parameter Name could not be found
NO ACCESS
User does not have write access
OUT OF BOUNDS
New value is not within the allowable range
NON NUMERIC
New value is not a number
PARAMETER DESCRIPTION
OPTIONS
P007
Gage Resolution
1 = .001", .02mm, 2 = .0005", .01mm,
3 = .0001", .002mm, 4 = .00005", .001mm
P011
Measurement Mode
0=ABS, 1=TIR, 2=INCR
P013
Units
1=Inches, 2=Millimeters
P019
Travel Reverse
0=Normal, 1=Reverse
P021
Hold On
0=Off, 1=On
P022
Hold Mode
0=Frz, 1=Min, 2=Max
P040
Ratio On
0=Off, 1=On
P063
Suppress Output
0=None, 1=Disable error and status
responses