Sefam S.Box
Maintenance
49
Engage the elbow seal on the pressure/flow sensor, while positioning the 3 ribs at the bottom
of the rear cover on the bottom cover.
Screw with a T6 tip equipped screwdriver, until it locks.
Connect the pressure / flow sensor and humidifier cables to the main board (sensor cable
passing in front of the humidifier cable).
Place the main board on the bottom cover. Align the back of the main board with the rear
cover, while positioning the pressure/flow sensor cable between the 2 capacitors.
Place the connector board in its back cover slot (insert the connectors in their slots).
Screw
Humidifier
connection
Sensor
connection
Humidifier cable in
front
of
sensor
cable
Sensor
cable
passage
Содержание Sefam S.Box
Страница 1: ......