10
Information and settings
Sefam S.Box
the maximum value of the ramp time adjustable by the patient when
the ramp is activated.
3G:
wireless mobile telecommunication technology for connecting
electronic devices.
Wi-Fi:
high-speed wireless transmission technology for connecting electronic
devices to one another.
Settings ranges
The following tables indicate the operating mode of the device, the default value, the minimum and
maximum values that can be set for each parameter, and the available functions.
Operating mode:
Parameter
Display
By default
Operating mode
CPAP
APAP
(Auto-CPAP)
APAP
(Auto-CPAP)
Adjustable parameters:
Parameter
Minimum
value
Maximum
value
Value by
default
Increment
Humidification level (if the
humidifier is installed)
OFF
10
02
01
Heating power of the heated
tube (if installed)
OFF
05
AUTO (with humid.)
01 (without humid.)
01
Mask select: theoretical leak
in the mask at 12.0 cm H
2
O
20 l/min
(Type:A)
60 l/min
30 l/min
(Type:B)
2 l/min
Ramp time
(if T Ramp is selected)
OFF
Maximum
ramp time
15 min
5 min
Maximum ramp time
(if T Ramp is selected)
OFF
45 min
45 min
5 min
Brightness of the display
01
10
05
01
In CPAP mode:
Set pressure
4 cm H
2
O
20 cm H
2
O
8 cm H
2
O
0.5 cm H
2
O
Ramp pressure (ramp start
pressure)
4 cm H
2
O
Prescribed pressure
4 cm H
2
O
0.5 cm H
2
O
In Auto-CPAP mode:
Ramp pressure (ramp start
pressure)
4 cm H
2
O
Maximum
set pressure
4 cm H
2
O
0.5 cm H
2
O
Maximum pressure
4 cm H
2
O or
minimum pressure
20 cm H
2
O
20 cm H
2
O
0.5 cm H
2
O
Minimum pressure
4 cm H
2
O
20 cm H
2
O or
maximum pressure
4 cm H
2
O
0.5 cm H
2
O
The device is switched on by holding down the start / standby button
. If the heated humidifier is
installed, it switches on automatically if the humidification level is set between 01 to 10.
Содержание Sefam S.Box
Страница 1: ......