Troubleshooting tips
Sefam S.Box
Troubleshooting tips
Device Messages
Message on the
display
Possible cause
Proposed solution
The mask is
disconnected.
Check the connection between the mask, the
breathing tube and the device. This message
disappears once you breathe into your properly
reconnected mask or press the
or
button.
This symbol
flashes on screen.
The device detected an
operating error in the
heated humidifier.
The device is operating without the heated
humidification function.
The corresponding error code is registered in
the memory of the machine and is accessible
by using the SEFAM Analyze software.
Refer to the section "Error messages".
This symbol
flashes on screen.
The device has detected a
malfunction of the S.Box
heated tube with ATC.
The device is operating without using the
function of the heated tube.
Check the tube connections to the device.
Unplug the device from all power sources. Plug
it in again and turn it on.
If the problem persists, contact Technical
Support Service.
The symbol
flashes quickly on
the status bar
display.
The SD card is either not
inserted or not properly
inserted in the device.
The SD card is locked.
The SD card is 90% full, or
more.
SD card error.
Insert the SD card properly into the SD card
slot.
Unlock the SD card and reinsert it into the
SD card slot.
Delete files on the SD card and give it back to
the patient.
Give a new SD card to the patient.
The error code
flashes.
(
XX = 2 figures)
.
The unit detected an
operating error.
Refer to the section "Error Messages".