Figure 7.8 Out Of Range (OOR) Correction
If no values have been selected for the Decimal Value (DV), no correction should be
selected. If Unit Values (UV) ranges were specified, the user has the option to change
the out of range measurements to the last value that was recorded before the out of
range measurement was recorded, or have it according to the max/min unit value as
specified earlier. In this case the software will know if the value is under the minimum
or maximum limit. After the correction has been made, the data needs to be
reconverted under File Reconvert Data, and data will be changed accordingly.
7.8
Time Synchronize Data (TSD)
Choose the Settings menu, TSD definition command. A dialog box appears:
Figure 7.9 Time Synchronous Data Settings
If the user wants to time synchronize two or more data files, it is possible to select
three options: 1 hour, 10 minutes, or 5 minutes. The restriction is that the sampling
interval selected must be greater than the initial interval, and it must also be dividable