Programming Commands 3-3
DS335 Synthesized Function Generator
parameters are separated by commas. Parameters shown in {} are optional
or may be queried while those not in {} are required. Commands that may be
queried have a question mark in parentheses (?) after the mnemonic.
Commands that may ONLY be queried have a ? after the mnemonic.
Commands that MAY NOT be queried have no ?. Do not send ( ) or { } as
part of the command.
All variables may be expressed in integer, floating point or exponential
formats (i.e., the number five can be either 5, 5.0, or .5E1). The variables i
and j usually take integer values, while the variable x take real number
values.
Function Output Control Commands
AECL
The AECL command sets the output to the ECL levels of 1 V peak-to-peak
with a -1.3 V offset. That is, from -1.8V to -0.8V.
AMPL (?) x
The AMPL command sets the output amplitude to x. The value x must
consist of the numerical value and a units indicator. The units may be VP
(Vpp) or VR (Vrms). For example, the command AMPL 1.00VR will set the
output to 1.0 Vrms. Note that the peak AC voltage (Vpp/2) plus the DC offset
voltage must be less than 5 Volts (for 50
Ω
source). Setting the amplitude to
0 Volts will produce a DC only (no AC function) output controlled by the
OFFS command.
The AMPL? query will return the amplitude in the currently displayed units.
For example, if the display is 3.0 Vrms the AMPL? query will return 3.0VR. If
a units indicator is sent with the AMPL? query (such as, AMPL? VP) the
displayed units will be changed to match the units indicator and the
amplitude returned in those units.
ATTL
The ATTL command sets the TTL output levels of 5V peak-to-peak with a
2.5V offset. That is, from 0V to 5V.
FREQ (?) x
The FREQ command sets the output frequency to x Hertz. The FREQ?
query returns the current output frequency. The frequency is set and
returned with 1
µ
Hz resolution. If the current waveform is NOISE an error will
be generated and the frequency will not be changed.
FUNC (?) i
The FUNC command sets the output function type to i. The correspondence
of i and function type is shown in the table below. If the currently selected
frequency is incompatible with the selected function an error will be
generated and the frequency will be set to the maximum allowed for the new
function. The FUNC? query returns the current function.
i
Function
0
SINE
1
SQUARE
2
TRIANGLE
3
RAMP
4
NOISE
INVT (?) i
The INVT command turns output inversion on (i=1) and off (i=0). The INVT?
query returns the current inversion status. This function is used with the
ramp waveform to set it for positive or negative slope.
Содержание DS335
Страница 2: ...DS335 Synthesized Function Generator...
Страница 6: ...iv SRS Symbols DS335 Synthesized Function Generator...
Страница 10: ...viii Specifications DS335 Synthesized Function Generator...
Страница 18: ...2 4 Introduction DS335 Synthesized Function Generator...
Страница 22: ...2 8 Features DS335 Synthesized Function Generator...
Страница 26: ...2 12 Function Setting DS335 Synthesized Function Generator...
Страница 29: ...Sweeps FSK 2 15 DS335 Synthesized Function Generator External Frequency Shift Keying FSK Example...
Страница 30: ...2 16 Sweeps FSK DS335 Synthesized Function Generator...
Страница 42: ...3 10 Programming Commands DS335 Synthesized Function Generator...
Страница 46: ...3 14 Programming Examples DS335 Synthesized Function Generator...
Страница 50: ...4 4 Troubleshooting DS335 Synthesized Function Generator...
Страница 74: ...5 8 Circuitry DS335 Synthesized Function Generator...