vi
General Information
Specifications
All performance specifications after 2 hour warm-up at 23
◦
C
±
1
◦
C
ambient, unless otherwise specified.
Output
Range
Full
Resolution
Max
Time
Settling
scale
(1 ppm)
current
const.
†
time
‡
V
μ
V
mA
μ
s
ms
1 V
±
1.010 000
1
50
160
50
10 V
±
10.100 00
10
50
160
50
100 V
±
101.000 0
100
25
400
100
†
Time required to settle to 63 % of final value
‡
Time to settle to 99.999 % of final value
Settling times measured for full-scale step into 10 M
Ω
load.
Accuracy and stability
Range
Accuracy
Stability
(23
◦
C
±
5
◦
C)
(
±
1
◦
C)
24 h
90 day
∗
1 year
∗
24 h
±
( ppm setting
+
μ
V)
1 V
7
+
2
12
+
6
25
+
10
1
+
1
10 V
7
+
6
12
+
15
25
+
20
1
+
3
100 V
8
+
60
12
+
80
25
+
100
1
+
20
∗
preliminary
Noise and drift
Range
Temperature
Noise
coe
ffi
cient
(typical)
0
◦
C – 40
◦
C
(0.1 – 10) Hz
10 Hz – 100 kHz
±
( ppm setting
+
μ
V)
/
◦
C
μ
Vrms
1 V
1
+
1
0.5
9
10 V
1
+
2
1.5
12
100 V
1
+
15
12
50
DC205 Precision DC Voltage Source