Protege IPG Clinician’s Manual
27
CAUTION:
Do not bring the suture needle in contact with an IPG, lead, or
extension, or the component may be damaged.
3. To stabilize the IPG within the pocket, pass a suture through the hole at the
top of the IPG header and secure it to connective tissue.
4. Check the entire system by fluoroscopy prior to closing to ensure proper
positioning of the lead or leads and that it is straight, with no sharp bends
or kinks.
5. Connect the communication wand to the patient programmer, place the
wand in a sterile bag, and position the wand over the IPG site.
6. Ensure that the patient programmer achieves effective communication with
the IPG and that the system is operational.
NOTE:
IPG output may not be identical to that of the trial stimulator at the same settings.