Protege IPG Clinician’s Manual
29
Replacing the IPG
The following steps outline the suggested procedure to replace an IPG:
1. Turn off the IPG or verify that it is turned off.
CAUTION:
Exercise care when using sharp instruments or electrocautery
around leads or extensions, or they may be damaged.
2. Open the IPG implant site per normal surgical procedure.
3. Insert the torque wrench through the septum of the IPG header and loosen
the setscrew by turning it counterclockwise.
CAUTION:
When performing the following step, do not bend the lead or
extension sharply; or it may be damaged.
4. Gently remove the lead or extension from the IPG header; then clean and dry
all connections, ensuring they are free of fluid and tissue.