UGA Series Universal Gas Analyzers
UGA Error Messages
4 –
75
Reserved
76
“IG voltage”,
() No power on the ion gauge controller board.
(2) The selected IG filament is burnt.
(3) The ion gauge failed to maintain the grid voltage. This
condition turns off the ion gauge.
77
“IG emission”,
The ion gauge failed to maintain the emission current. This
condition turns off the ion gauge.
78
“IG off failed”,
The ion gauge is not turned off in time, probably due to
miscommunication between the main control board
and the auxiliary control board.
79
Reserved
80
“RGA off failed”,
The RGA is not turned off in time, probably due to
miscommunication between the main control board and
the auxiliary control board.
8
“No Mux detected”,
Multi-inlet valve controller is not detected.
82
“Mux malfunction”,
The multi-inlet valve channel reported from the controller
does not match with that of the main control board,
probably due to miscommunication between the main
control board and the auxiliary control board.
83 - 84 Reserved
85
“No Vent valve”,
No vent valve was detected.
86 – 88 Reserved