3.4 Commands
3 – 23
Command Error
LCME?
Query the last command error code. Valid codes are:
Value
Definition
0
No execution error since last
LCME?
1
Illegal command
2
Undefined command
3
Illegal query
4
Illegal set
5
Missing parameter(s)
6
Extra parameter(s)
7
Null parameter(s)
8
Parameter bu
ff
er overflow
9
Bad floating-point
10
Bad integer
11
Bad integer token
12
Bad token value
13
Bad hex block
14
Unknown token
*IDN
Example:
LCME?
4
The error (4, “Illegal set”) is due to the missing “
?
”.
Last Button
LBTN?
Query the last button that was pressed. The values returned are:
Value
Button
0
no button pressed since last
LBTN?
1
[Setpoint]
2
[Output]
3
[Ramp Start
/
Stop]
4
[Shift]
5
[Select]
6
[On
/
O
ff
]
7
[ ]
/
[ ]
8
[ ]
/
[ ]
LBTN?
Example:
5
SIM960 Analog PID Controller