![SRS Labs SIM960 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/srs-labs/sim960/sim960_operation-and-service-manual_1354206044.webp)
3 – 18
Remote Operation
Status Byte
*STB? [
i
]
Reads the Status Byte register [bit
i
].
Execution of the
*STB?
query (without the optional bit
i
) always
causes the
−
STATUS signal to be deasserted. Note that
*STB?
i
will
not
clear
−
STATUS, even if bit
i
is the only bit presently causing the
−
STATUS signal.
*STB?
Example:
16
Service Request Enable
*SRE(?) [
i,
]
{
j
}
Set (query) the Service Request Enable register [bit
i
]
{
to
j
}
.
*SRE 32; *SRE? 5
Example:
1
Standard Event Status
*ESR? [
i
]
Reads the Standard Event Status Register [bit
i
].
Upon executing
*ESR?
, the returned bit(s) of the ESR register are
cleared.
GAIN 0
Example:
ESR?
16
The binary value (16) corresponds to an Execution Error, since
GAIN
0
is an illegal value (minimum gain is 0.1).
Standard Event Status Enable
*ESE(?) [
i,
]
{
j
}
Set (query) the Standard Event Status Enable Register [bit
i
]
{
to
j
}
.
*ESE 16
Example:
Comm Error Status
CESR? [
i
]
Query Comm Error Status Register [for bit
i
].
Upon executing a
CESR?
query, the returned bit(s) of the CESR
register are cleared.
CESR?
Example:
0