CALIBRATION cont’d 

 

SETTING CALIBRATION 

 

NOTE:

  Ensure that nothing is in contact with the scale system during this procedure.  Remove 

hands from the system when noting the displayed calibration results.   
 

STEP 1:

  Access the Calibration Switch by breaking the Calibration Seal on the underside of the 

display  extrusion.  Put  the  scale  system  into  the  Calibration  Mode  by  switching  the  calibration 
switch on the display board (Error! Reference source not found.).  

NOTE:  USE A NON-METALLIC DEVICE TO SLIDE THE SWITCH.  FAILURE TO DO 

SO WILL RESULT IN DAMAGE TO THE BOARD

 

STEP  2:

   

Select  known  calibrated  weights,  traceable  to  NIST,  up  to  the  Full  Scale  value 

(maximum capacity).   

STEP 3:

  Press the “

MENU”

 button until “

1000.0Lb FULL

” is displayed.  Set the 

FULL

 value to the 

actual quantity of calibrated weight being used for Full Scale.  Use the “

UP”

 arrow button to select the 

digit  and  the  “

DOWN”

  arrow  button  to  change  digit.    Press  “

ENTER

”  to  save  changes.  Press 

WEIGH

” button to abort any changes.   

STEP 4:

  Press the 

MENU

 button until “

500.0Lb HALF

” is displayed.   

NOTE

:    The  Half-Scale  value  is  a  value  between  zero  and  the  Full  Scale  values.    It  is  usually 

close to half the Full Scale value. 

Set the 

HALF

 value to the actual quantity of calibrated weight being used for Half Scale.  Use the 

UP”

 arrow button to  select  the digit and the “

DOWN”

  arrow button to change the digit. 

 Press 

“ENTER”

 to save changes. Press “

WEIGH

” to abort the changes.  

STEP 5:

  Press 

MENU

 button until “

3 PT CAL

” is displayed.  Press the “

UP”

 arrow button. 

STEP 6:  

Display will read “

ZERO

”.  Ensure that all weight has been removed from the platform 

and press the “

UP” 

arrow button. 

STEP 7:

   Display will read “

ADD HALF

”.  Place the calibrated weight on platform for 

HALF 

Scale.  Allow weight to stabilize.  Press “

UP”

 arrow button to save change. 

STEP  8:

    Display  will  read  “

ADD  FULL

”.  Place the calibrated weight  on platform for 

FULL 

Scale.  Allow weight to stabilize.  Press “

UP” 

arrow to save change.   

STEP  9

:    Press 

“ENTER” 

button  to  save  the  calibration,  or  “

WEIGH

”  button

 

to  exit  without 

saving

.  

 

STEP  10:

    Switch  the  scale  system  out  of  the  Calibration  Mode  on  the  display  board  (Error! 

Reference source not found.).   

NOTE

:  Ensure that no wires have become unplugged during calibration 

STEP  11:

    Calibration  Seal  must  be  placed  over  the  access  hole  on  the  display  extrusion  by 

authorized personnel only. 

Содержание SR463iR-3H

Страница 1: ...Part No MAN463iR H_201118 Page 1 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments Inc In Floor Platform Scale Operating and Service Manual SR463iR H SERIES ...

Страница 2: ...ECKLIST SR463IR 4H SINGLE SHIPMENT 5 ASSEMBLY 6 REPLACEMENT PARTS AND ACCESSORIES 10 SYSTEM DESCRIPTION 10 INTENDED USE 10 CLEANING AND DISINFECTING 11 STORAGE AND TRANSPORTATION 12 SPECIFICATIONS 12 BUTTON FUNCTIONS 13 BASIC SYSTEM OPERATION 14 THEORY OF OPERATION 15 CALIBRATION 16 TROUBLESHOOTING 18 WARRANTY 19 TABLE OF FIGURES Figure 1 Wall Box Assembly Conduit Placement 6 Figure 2 Conduit Plac...

Страница 3: ...SSEMBLY 1 ea PIT FRAME ASSEMBLY 1 ea CONDUIT 1 ea LONG FLAT HEAD SCREW 4 ea INSTALLATION INSTRUCTIONS 1 ea 2nd Shipment CABLE ASSEMBLY 1 ea PLATFORM ASSEMBLY 32 IN X 36 IN 1 ea FRONT PANEL ASSEMBLY 1 ea HANDRAIL 1 ea LONG FLAT HEAD SCREW 8 ea LONG BUTTON HEAD SCREW 4 ea STAR WASHERS 4 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea MANUAL 1 ea DESCRIPTION QUANTITY Box 1 FRONT PANEL WALL BOX ASS...

Страница 4: ...ASSEMBLY 1 ea PIT FRAME ASSEMBLY 1 ea CONDUIT 1 ea LONG FLAT HEAD SCREW 4 ea INSTALLATION INSTRUCTIONS 1 ea 2nd Shipment CABLE ASSEMBLY 1 ea PLATFORM ASSEMBLY 32 IN X 50 IN 1 ea FRONT PANEL ASSEMBLY 1 ea HANDRAIL 1 ea LONG FLAT HEAD SCREW 8 ea LONG BUTTON HEAD SCREW 4 ea STAR WASHERS 4 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea MANUAL 1 ea DESCRIPTION QUANTITY Box 1 FRONT PANEL WALL BOX AS...

Страница 5: ...ASSEMBLY 1 ea PIT FRAME ASSEMBLY 1 ea CONDUIT 1 ea LONG FLAT HEAD SCREW 4 ea INSTALLATION INSTRUCTIONS 1 ea 2nd Shipment CABLE ASSEMBLY 1 ea PLATFORM ASSEMBLY 48 IN X 72 IN 1 ea FRONT PANEL ASSEMBLY 1 ea HANDRAIL 1 ea LONG FLAT HEAD SCREW 8 ea LONG BUTTON HEAD SCREW 4 ea STAR WASHERS 4 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea MANUAL 1 ea DESCRIPTION QUANTITY Box 1 FRONT PANEL WALL BOX AS...

Страница 6: ... Area Installation Notes Pit Area Installation Notes This is a rectangular frame so it is important to check the placement specifications to make sure the conduit connection is properly placed facing the wall where the conduit cable connection is located There are holes cut for conduit entry in the center of two sides of the frame so the scale can be placed with either the long or short side again...

Страница 7: ...and cable should be left free so cable can be easily connected to the Platform Assembly STEP 7 Figure 5 When the concrete has set remove and discard the installation tabs and screws Install the four long flat head screws into holes left by plate removal Figure 4 Concrete Fill and Installation Tabs Figure 5 Hardware Placement Figure 6 Cable Connection Inside Frame STEP 8 Remove the Platform Assembl...

Страница 8: ...Adjustment Bolts so that the Platform Assembly is at an equal distance from all sides of the Pit Frame Assembly Tighten bolts to secure Platform Assembly in place STEP 11 Figure 5 Set the Platform Deck onto the platform base and secure in place with the four 4 top mounting screws Allow 1 8 gap between Platform Assembly and floor STEP 12 Install building ground wire and green ground wire from Displ...

Страница 9: ... cont d STEP 15 Install Front Panel Assembly Figure 8 to the previously mounted Wall Box Assembly Use supplied star washers and screws Ensure the screws are securely tightened STEP 16 Restore power to the display STEP 17 Secure the handrail to the platform deck using the eight 8 included flat head screws Figure 8 Front Panel Assembly Figure 9 Handrail Attached to Scale WARNING HAND RAIL IS INTENDE...

Страница 10: ...Four 4 identically matched load cell transducers are strategically placed to ensure an accurate representation of the patient s weight regardless of weight distribution The SR463iR H Series display units use 120v AC line power that is mounted in a recessed wall box so the display is flush with the wall Figure 8 The patient s weight is displayed on a 16 character dot matrix liquid crystal display T...

Страница 11: ...e a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants Disinfectant solutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with ...

Страница 12: ... in 81 cm x 91cm SR463iR 3H 32 in x 50 in 81 cm x 127 cm SR463iR 4H 48 in x 72 in 122 cm x 183 cm DISPLAY TYPE 16 Character dot matrix LCD DISPLAY RESOLUTION 0 1 lb 0 1 kg ACCURACY 0 1 1 digit of displayed resolution for calibrated range ZERO One button Auto Tare AUTO POWER DOWN Adjustable between 30 and 300 seconds AVERAGING Automatic digital filter POWER SUPPLY 120v AC 50 60 Hz CALIBRATION Calib...

Страница 13: ...on operational on this model RECALL Press to recall the last stored weight The stored weight will be erased when the scale is zero or another stable weight is stored BMI Press to calculate BMI When the BMI is pressed the default value HT 65 in or HT 165 cm is displayed If there is no stored stable weight the display will indicate NO WEIGHT DATA and then go back to the weigh screen WT 0 0 Lb MENU P...

Страница 14: ...time or to change the value of a selected digit when setting time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Make sure the scale is free and clear of any obstructions Press and hold the ZERO WEIGH button The displayed message will indicate HOLD TO ZERO and count down to zero Release the button when display message indicates HANDS OFF Make sure that nothing is in contact with the scale whi...

Страница 15: ...digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the reference voltage applied to the converter The residual voltage on the integrating capacitor is then multiplied by a factor and again discharged at a rate proportional to the reference...

Страница 16: ...USE barbells or un calibrated weights STEP 2 Zero the scale by pressing and holding the ZERO WEIGH button STEP 3 Place the calibrated weight on the scale Wait for scale to stabilize note scale reading STEP 4 Scale readings should be within Calibration Tolerance Table Figure 11 Figure 11 Calibration Tolerance Table Figure 12 Calibration Switch Diagram IMPORTANT CALIBRATION Qualified service personn...

Страница 17: ...ange digit Press ENTER to save changes Press WEIGH button to abort any changes STEP 4 Press the MENU button until 500 0Lb HALF is displayed NOTE The Half Scale value is a value between zero and the Full Scale values It is usually close to half the Full Scale value Set the HALF value to the actual quantity of calibrated weight being used for Half Scale Use the UP arrow button to select the digit an...

Страница 18: ...ers only appear on half of the display Press the WEIGH button Display shows no reading at all Check to ensure that circuit breaker to the scale is on Display reads WEIGHT OVERLOAD Disconnect the transducer cable from the platform If the WEIGHT OVERLOAD message goes away the problem is in the platform Check to make sure the platform transducer wires are intact and pit is clean and dry For additiona...

Страница 19: ...r obligations or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages Any misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes ...

Страница 20: ... Service Manual Part No MAN463iR H_201118 Page 20 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S S Instruments Inc Precision Technology in Perfect Balance ...

Отзывы: