Model SR411
i
Patient Floor Scale
Operating and Service Manual
Part No. MAN411
i
_MICROBAN_200610
Page 3 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2020
S
Instruments, Inc.
PACKING CHECKLIST – Model SR411
i
Patent Floor Scale
ASSEMBLY
STEP 1
: Unpack the scale system and check parts against the
PACKING CHECKLIST
. If
there are any missing or damaged parts, call the Service Hotline at: 1-800-654-6360.
STEP 2
: Determine installation location for scale and display, keeping in mind the display cable
length is six (6) feet and the display power cord length is four (4) feet. The display may lay
flat on a desk or end table. If desired, provisions for wall mount have been incorporated, and
can be accomplished in the manner described below.
A.
The display should be installed by qualified personnel. The mounting hardware provided
will cover most wall applications.
B.
Determine a position on the wall comfortable to the user. Mark the location of the two
(2) mounting holes on the wall (holes are spaced 2.5” on center apart) using the template
provided on page 15.
C.
Drill the two (2) pilot holes, no larger than ¼” diameter in the locations marked. Install
the two (2) self-drilling screw anchors.
D.
Install the two (2) #8 x 1” flat head screws supplied, leaving approximately ¼”
projecting from the wall.
STEP 3
: Place the SR411
i
scale system on a flat, hard floor surface. Ensure all four corners of
scale are in contact with the floor and scale does not rock
.
NOTE:
Scale should not be placed on carpet or uneven floor surfaces as this may interfere
with scale accuracy.
STEP 4
: Connect the display cable to the scale, then connect the power cable to the display and
plug power adapter into wall.
DESCRIPTION
QUANTITY
SCALE BASE
1 ea.
SCALE DISPLAY
1 ea.
MOUNTING HARDWARE
1 ea.
AC WALL POWER ADAPTER (MEDICAL GRADE)
1 ea.
CERTIFICATE OF CALIBRATION
1 ea.
MANUAL
1 ea.