CALIBRATION (Cont’d) 

 

Auto Power Down Timer: (Optional)

 

a)

 

Select the Auto Power Down Timer from the CAL menu by pressing the Left Button 
when the display shows “

sec

”. 

b)

 

The display will show four digits <0123>. 

c)

 

Set  the  desired  time  in  seconds  by  selecting  the  first  digit  and  toggle  through  the 
digits using the Right Button until desired value is shown. Press the Left Button and 
continue  the  same  with  the  remaining  digits.  The  value  cannot  be  less  than  15  or 
greater than 900. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

 

Gravitational Acceleration: (Optional)

 

NOTE

:  Setting  the  local  gravitational  acceleration,  “

gLOC

”,  will  not  affect  the 

calibration. The local gravitational acceleration, “

gLOC

”, will become the calibrated 

gravitational acceleration, “

gCAL

 

only when it is recalibrated. 

a)

 

Select  the  Gravitational  Acceleration  from  the  CAL  menu  by  pressing  the  Left 
Button when the display shows “

grav

”. 

b)

 

The  display  will  show  only  the  four  digits  to  the  right  of  the  decimal  point 
<9.XXXX>. 

c)

 

Set the local  gravitational acceleration in m/s

2

 by selecting the first digit and toggle 

through the digits using the Right Button until desired value is shown. Press the Left 
Button and continue the same with the remaining digits. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

 

Full-Scale value: (Optional)

 

NOTE

: The Full-Scale value should be at least 200 kilograms (440 pounds) (half the 

maximum scale capacity) 

a)

 

Select  the  Full-Scale  from  the  CAL  menu  by  pressing  the  Left  Button  when  the 
display shows “

FULL

”. 

b)

 

The  display  will  show  only  the  five  digits  <XXXX.X>  with  the  thousand-place 
flashing <1>. Select past to the next digit by pressing the Left Button. 

c)

 

Once  in  the  hundreds-place,  set  the  full  scale  value  (maximum  load)  by  toggling 
using  the  Right  Button  through  the  digits  until  desired  value  is  shown.  Select  the 
desired value by pressing the Left Button and continue the same with the remaining 
digits. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

 

Содержание SR411i

Страница 1: ...ICROBAN_200610 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments Inc Operating and Service Manual Model SR411i Patient Floor Scale ...

Страница 2: ... Inc TABLE OF CONTENTS PACKING CHECKLIST MODEL SR411I 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 MICROBAN ANTIMICROBIAL PROTECTION 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUNCTIONS 7 BASIC SYSTEM OPERATION 8 THEORY OF OPERATION 9 CALIBRATION 10 TROUBLESHOOTING 13 WARRANTY 14 DISPLAY MOUNTING TEMPLATE 15 TABLE OF FIGURES Figure 1 SR411i Button Labe...

Страница 3: ...ed by qualified personnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are spaced 2 5 on center apart using the template provided on page 15 C Drill the two 2 pilot holes no larger than diameter in the locations marked Install the two 2 self drilling screw ...

Страница 4: ...ograms INTENDED USE The SR411i Patient Floor Scale is a unique lightweight low profile stand on scale system for weighing ambulatory patients The wide platform is designed to easily accommodate larger patients The SR411i Patient Floor Scale is a preferred means of gathering patient weight data of patients weighing up to 880 pounds 400 kilograms MICROBAN ANTIMICROBIAL PROTECTION Microban antimicrob...

Страница 5: ...areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants Disinfectant solutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampen...

Страница 6: ...lution for calibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPPLY Medical Grade AC Wall Power Supply CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS Normal operating conditions for this product Ambient Temperature Range 68 F ...

Страница 7: ...atient is in position on the scale the weight will be displayed NOTE When the system is zeroed the Hold Recall memory is erased LB KG BUTTON Patient weight data may be viewed in either pounds or kilograms Pressing the lb kg button allows the operator to toggle between the two readings HOLD RECALL BUTTON Scale will automatically lock a stable weight that is greater than five pounds and MEM will be ...

Страница 8: ...ternatively press the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the patient on the platform Weight will be displayed in a few seconds A stable weight will automatically be stored in memory STEP 3 To manually hold a weight reading at any time during the weighing process press the Hold Recall button NOTE When the ...

Страница 9: ...d by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the reference voltage applied to the converter The residual voltage on the integrating capacitor is then multiplie...

Страница 10: ...fied service personnel should perform this procedure Load cells have no user serviceable components and should not be tampered with for any reason Re calibration is generally not required but should be verified periodically to ensure accuracy The recommendation for calibration check is at least once every 12 months or as individual maintenance policy requires CAUTION The integrated circuits and se...

Страница 11: ...that nothing is in contact with the scale system during this procedure Remove hands from the system when noting the displayed calibration results CAL MENU data Displays saved calibration and model information unit Set the available units lb only kg only or selection of lb or kg with mode button sec Set Auto Power Down Timer between 15 and 900 seconds grav Set Gravitational Acceleration in m s2 Ful...

Страница 12: ...leration from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitational acceleration in m s2 by selecting the first digit and toggle through the digits using the Right Button until desired value is shown Press the Left Button and continue the same with the remaining digits...

Страница 13: ...aceable to NIST DO NOT USE barbells or uncalibrated weights The displayed calibrated weight should be within the Accuracy tolerance as listed in the SPECIFICATIONS table on page 5 a Select Two point calibration curve by pressing the Left Button when the display shows 2Pt b With no load applied zero the scale by pressing the Left Button when the display shows ZERO c Display will show FULL apply the...

Страница 14: ...r liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages Any misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes to purchaser up...

Страница 15: ...ay flat on a desk or end table If desired provisions for wall mount have been incorporated and can be accomplished in the manner described below A The display should be installed by qualified personnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are spaced...

Страница 16: ...o MAN411i_MICROBAN_200610 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S By SInstruments Inc Precision Technology in Perfect Balance ...

Отзывы: