BUTTON FUNCTIONS
Figure 1: SR411
i
Button Label
ZERO / WEIGH
The “
Zero/Weigh
” button is used to zero the system before placing a patient onto the scale
system. Ensure that nothing is in contact with the weighing surface during this procedure.
Press and hold the “
Zero/Weigh
” button until the display reads “
ZERO
”, then release and
wait until the display reads
“0.0”
. When the patient is in position on the scale, the weight will be
displayed.
NOTE
: When the system is zeroed, the Hold/Recall memory is erased.
LB / KG BUTTON
Patient weight data may be viewed in either pounds or kilograms. Pressing the “
lb/kg
” button
allows the operator to toggle between the two readings.
HOLD/RECALL BUTTON
Scale will automatically lock a stable weight that is greater than five pounds and “
MEM
” will
be displayed in the upper left of the display. To manually hold a weight reading at any time
during the weighing process, press the “
Hold/Recall
” button. To recall a weight in memory,
press the “
Hold/Recall
” button after the display has powered down or with less than five
pounds on the scale.
NOTE:
When the scale is off, pressing any of the buttons will turn on the scale. Press and
hold the
“Zero/Weigh”
button to erase the last held weight from memory.