TROUBLESHOOTING 

 

 

SYMPTOM 

REASON/CORRECTIVE ACTION 

The characters only appear on half of the 
display. 

Press the

 “WEIGH” 

button or remove one battery.  

Wait five seconds, then re-install the

 

battery and try

 

the

 “WEIGH” 

button again.

   

The display lights appear to work, but do 
not respond to button activation. 

Check to ensure the membrane switch label is correctly 
plugged into display board. 
Check  to  ensure  the  calibration  switch  is  not  in  the 
Calibration Mode (Figure 12). 

The display shows no reading at all. 

Check  to  ensure  batteries  are  installed  correctly 

(see 

directions 

for 

BATTERY  REPLACEMENT

).  

Check  display  cable  to  make  sure  it  is  connected 
securely. 

The display feels loose when raising or 
lowering. 

Remove  the  black  plastic  plug  located  below  the 
display rotation assembly. Insert a 5/32” Allen Key and 
slightly tighten the tensioning screw until the mast feels 
“tight” again. 

For additional information or assistance, telephone our Service Hotline: 1-800-654-6360 

or e-mail: 

[email protected]

 

Содержание SR Scales SR725i

Страница 1: ...AN_200603 Page 1 of 24 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S In...

Страница 2: ...N 9 STORAGE AND TRANSPORTATION 10 CLEANING AND DISINFECTING 11 SPECIFICATIONS 12 BUTTON FUNCTIONS 13 BASIC SYSTEM OPERATION 14 BATTERY REPLACEMENT 16 THEORY OF OPERATION 17 CALIBRATION 18 CONFIGURING...

Страница 3: ...20 SInstruments Inc PACKING CHECKLIST Model SR725i Wheelchair Scale System DESCRIPTION QUANTITY Box 1 SCALE ASSEMBLY 32 in x 30 in 81 cm x 76 cm 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea CALIBRATION...

Страница 4: ...0 SInstruments Inc PACKING CHECKLIST Model SR725i L Wheelchair Scale System DESCRIPTION QUANTITY Box 1 SCALE ASSEMBLY 36 in x 30 in 91 cm x 76 cm 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea CALIBRATIO...

Страница 5: ...Diagram Figure 2 Parts Identification List STEP 3 Figure 1 Lay the scale on the floor with the feet on the floor STEP 4 Adjust leveling feet located in each of the four corners of the underside of the...

Страница 6: ...r bumper from the display base Figure 3 Rotate the display 180 degrees Re insert mounting screw and rubber bumper and tighten Note To provide correct alignment and to secure the display to the mast pr...

Страница 7: ...the top of the Mast Tighten the two 2 set screws to secure the Printer Mounting Collar to the Mast NOTE The three 3 holes on the Printer Mounting Collar should be towards the top of the Mast facing th...

Страница 8: ...om email sri srinstruments com Copyright 2020 SInstruments Inc REPLACEMENT PARTS and ACCESSORIES Part Description SR8219 Printer Kit optional contains all printer parts listed below FRAP1300 Printer F...

Страница 9: ...p to 400 hours of weight readings before needing replacement This eliminates the need for an external battery charger or the danger of an AC power supply cord on a portable scale The patient s weight...

Страница 10: ...ns should not vary outside the following conditions Relative Humidity 0 to 85 Ambient Temperature 14 F to 122 F 10 C to 50 C TRANSPORTATION To transport the scale pull up on the Snap Lock Figure 7 Fol...

Страница 11: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Страница 12: ...ge AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds ENTER Stores displayed reading in memory RECALL Recalls last stored stable weight height and BMI AV...

Страница 13: ...en stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and BMI Also indicated on display when pap...

Страница 14: ...selected digit when setting time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Make sure the scale is free and clear of any obstructions and press and hold the ZERO WEIGH button The displayed me...

Страница 15: ...5 in Use the UP or DOWN arrows to adjust the default height to the value of the patient s height NOTE To calculate the BMI a patient s stable weight needs be stored as indicated in WEIGHING above if n...

Страница 16: ...he panel screw on the Battery Compartment Cover 13 and remove the battery compartment cover STEP 3 Remove and replace ALL six 6 D cell batteries Refer to diagram in the battery compartment for placeme...

Страница 17: ...rs serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloaded...

Страница 18: ...second weight should be less than half of the first weight DO NOT USE barbells or un calibrated weights STEP 2 Zero the scale by pressing and holding ZERO WEIGH button STEP 3 Place the first calibrate...

Страница 19: ...f the first weight DO NOT USE barbells or uncalibrated weights STEP 3 Press the MENU button until FULL 1000 00 Lb is displayed Set the FULL value of the first selected calibrated weight Use the UP arr...

Страница 20: ...ING SEND BUTTON STEP 1 Press the MENU key until DATA OUT PRINT is displayed STEP 2 Press UP or DOWN key to select between EHR and PRINT STEP 3 Press the ENTER key to save EHR OUTPUT Electronic Medical...

Страница 21: ...k but do not respond to button activation Check to ensure the membrane switch label is correctly plugged into display board Check to ensure the calibration switch is not in the Calibration Mode Figure...

Страница 22: ...iabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no...

Страница 23: ...and Service Manual Part No MAN725i_MICROBAN_200603 Page 23 of 24 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments co...

Страница 24: ...MAN725i_MICROBAN_200603 Page 24 of 24 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Отзывы: