THEORY OF OPERATION 

 

SR  Instruments  patient  weighing  systems  are  digital  scales.  Strain-gauge  force  cells  convert  the 
force  of  an  applied  weight  into  an  analog  signal.  This  signal  is  amplified  by  an  operational 
amplifier  and  converted  to  a  digital  signal  by  an  on-chip  analog  to  digital  converter.  The  digital 
signal is filtered, converted to appropriate units, and displayed on the liquid crystal display. 
 
Strain-gauge  force  cells  each  contain  four  strain  gauges  mounted  in  a  full  Wheatstone-bridge 
configuration. These bridges convert the physical movement of the force cell, due to the applied 
mass  on  the  system,  into  minute  changes  in  electrical  resistance.  These  changes  in  resistance 
produce a voltage difference across the Wheatstone-bridge, which is amplified by the operational 
amplifier. The amplifier is configured to current sum the output of each cell, with potentiometers 
serving  to  normalize  the  sensitivity  (voltage  out  per  unit  of  weight  applied)  of  each  bridge.  The 
offset  potentiometer  produces  a  small  current,  which  nulls  the  output  of  the  amplifier  for  an 
unloaded system.  
 
The output of the operational amplifier is digitized by the analog to digital converter. The sigma-
delta  converter  sums  a  rapid  sequence  of  0's  (0  volts)  and  1's  (reference  voltage)  to  achieve 
balance with the input signal from the amplifier.  
 
The micro-controller filters the digital output of the analog to digital converter, subtracts the value 
saved during the system zero operation and scales the filtered output, and then displays the result 
on  the  liquid  crystal  display.  The  micro-controller  performs  a  moving-median  filter  of  data  for 
continuous  weigh  and,  for  AutoHold;  the  micro-controller  performs  checks  for  signal  stability 
before locking in on the reading. If the data variance is greater than 0.1% in the AutoHold mode, 
the micro-controller will reset the filter and start a new filtering period. 
 
The micro-controller can be placed in a calibration mode, where the system can be re-calibrated. 
In the calibration mode, the system slope is calculated from two points (zero and full scale) in the 
2-point  calibration  mode  or  the  slope  and  change  in  slope  is  calculated  from  three  points  (zero, 
half, and full scale) in the 3-point calibration mode. 

Содержание SR Scales SR725i

Страница 1: ...AN_200603 Page 1 of 24 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S In...

Страница 2: ...N 9 STORAGE AND TRANSPORTATION 10 CLEANING AND DISINFECTING 11 SPECIFICATIONS 12 BUTTON FUNCTIONS 13 BASIC SYSTEM OPERATION 14 BATTERY REPLACEMENT 16 THEORY OF OPERATION 17 CALIBRATION 18 CONFIGURING...

Страница 3: ...20 SInstruments Inc PACKING CHECKLIST Model SR725i Wheelchair Scale System DESCRIPTION QUANTITY Box 1 SCALE ASSEMBLY 32 in x 30 in 81 cm x 76 cm 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea CALIBRATION...

Страница 4: ...0 SInstruments Inc PACKING CHECKLIST Model SR725i L Wheelchair Scale System DESCRIPTION QUANTITY Box 1 SCALE ASSEMBLY 36 in x 30 in 91 cm x 76 cm 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea CALIBRATIO...

Страница 5: ...Diagram Figure 2 Parts Identification List STEP 3 Figure 1 Lay the scale on the floor with the feet on the floor STEP 4 Adjust leveling feet located in each of the four corners of the underside of the...

Страница 6: ...r bumper from the display base Figure 3 Rotate the display 180 degrees Re insert mounting screw and rubber bumper and tighten Note To provide correct alignment and to secure the display to the mast pr...

Страница 7: ...the top of the Mast Tighten the two 2 set screws to secure the Printer Mounting Collar to the Mast NOTE The three 3 holes on the Printer Mounting Collar should be towards the top of the Mast facing th...

Страница 8: ...om email sri srinstruments com Copyright 2020 SInstruments Inc REPLACEMENT PARTS and ACCESSORIES Part Description SR8219 Printer Kit optional contains all printer parts listed below FRAP1300 Printer F...

Страница 9: ...p to 400 hours of weight readings before needing replacement This eliminates the need for an external battery charger or the danger of an AC power supply cord on a portable scale The patient s weight...

Страница 10: ...ns should not vary outside the following conditions Relative Humidity 0 to 85 Ambient Temperature 14 F to 122 F 10 C to 50 C TRANSPORTATION To transport the scale pull up on the Snap Lock Figure 7 Fol...

Страница 11: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Страница 12: ...ge AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds ENTER Stores displayed reading in memory RECALL Recalls last stored stable weight height and BMI AV...

Страница 13: ...en stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and BMI Also indicated on display when pap...

Страница 14: ...selected digit when setting time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Make sure the scale is free and clear of any obstructions and press and hold the ZERO WEIGH button The displayed me...

Страница 15: ...5 in Use the UP or DOWN arrows to adjust the default height to the value of the patient s height NOTE To calculate the BMI a patient s stable weight needs be stored as indicated in WEIGHING above if n...

Страница 16: ...he panel screw on the Battery Compartment Cover 13 and remove the battery compartment cover STEP 3 Remove and replace ALL six 6 D cell batteries Refer to diagram in the battery compartment for placeme...

Страница 17: ...rs serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloaded...

Страница 18: ...second weight should be less than half of the first weight DO NOT USE barbells or un calibrated weights STEP 2 Zero the scale by pressing and holding ZERO WEIGH button STEP 3 Place the first calibrate...

Страница 19: ...f the first weight DO NOT USE barbells or uncalibrated weights STEP 3 Press the MENU button until FULL 1000 00 Lb is displayed Set the FULL value of the first selected calibrated weight Use the UP arr...

Страница 20: ...ING SEND BUTTON STEP 1 Press the MENU key until DATA OUT PRINT is displayed STEP 2 Press UP or DOWN key to select between EHR and PRINT STEP 3 Press the ENTER key to save EHR OUTPUT Electronic Medical...

Страница 21: ...k but do not respond to button activation Check to ensure the membrane switch label is correctly plugged into display board Check to ensure the calibration switch is not in the Calibration Mode Figure...

Страница 22: ...iabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no...

Страница 23: ...and Service Manual Part No MAN725i_MICROBAN_200603 Page 23 of 24 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments co...

Страница 24: ...MAN725i_MICROBAN_200603 Page 24 of 24 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Отзывы: