P/N 20126 Rev B
Page 6
Basic Care
1.
Spot is a precision instrument. DO NOT drop it or subject it to physical shock.
2.
Spot is NOT waterproof. If you accidentally drop it into liquid or liquid is spilled on the device, contact
PediaVision Support immediately. If minor water droplets are found on Spot, wipe off with a soft dry cloth.
3.
DO NOT leave Spot in excessive heat, such as in a car or in direct sunlight. High temperatures can cause
Spot to malfunction.
4.
Spot contains precision electronic circuitry. Never attempt to disassemble Spot; this may result in bodily
harm and will void any warranty.
5.
DO NOT use cleaners that contain organic solvents to clean Spot’s body or the front glass. Instead, use the
provided cleaning cloth and for stubborn stains use with a small amount of water or a non-abrasive
cleaner. Always keep the front glass free of fingerprints, dirt and dust.
6.
When plugging the AC power connector into Spot, DO NOT force the power connector into the Spot
device as this may cause damage and will not be covered under warranty.
7.
DO NOT store Spot in its case with the power cord plugged in, this could cause damage to both the power
cord and the Spot device. If plugged in, store in a location with no interference with the power cord.
8.
DO NOT use the wrist strap as a means to carry the device, it is not designed to carry the weight of Spot.
9.
DO NOT push or scratch the touchscreen display surface with anything stiff or rigid. A stylist is
acceptable for use with the device.
10.
DO NOT spill liquids on Spot. Please wipe off drops of liquid like saliva and water on LCD panel surface
immediately.
Warnings
1.
Changes or modifications not expressly approved by PediaVision could void the user’s authority to
operate the equipment.
2.
The supplied AC cord set should be used exclusively. Replacement or substitution may only be done with a
CSA or UL approved hospital-grade AC cord set.
3.
The supplied power supply/charger may not be substituted and shall only be replaced by an approved
unit supplied by PediaVision.
4.
Interconnection of this medical device with other medical devices, medical systems or non-medical
devices shall be evaluated to the requirements of CSA/IEC 60601-1 in the end use application.
5.
This equipment is ESD sensitive; do not touch connector pins during use.
6.
If the power cord or external power supply becomes damaged or its integrity is suspect, or the building
power receptacle outlet is ungrounded, then the Spot shall only be used with the internal battery power
until replacement parts become available.
7.
The operator shall not touch accessible conductive parts (ac/dc adaptor output connector, power input
connector and USB port) and the subject simultaneously.