Operating Manual

9

1300-87075-1A

User Interface

Plug in the system to turn the power on. Once the system is turned 

on, the Spex logo is displayed. The home screen appears as shown 

to the left. If the system is in ECO mode, it will prompt you to wake it.

This screen will show upon power up. It will instruct you to close the 

drawer to home the system and enable the unit to be ready to seal, 

simply push the drawer in.

 

The screen will then show the sealing time, heater set temperature 

and heater actual temperature along with the active status of the 

heater block, cooling or heating depending on the temperature.

 

This is the main sealing screen. Here, you will see the main button to 

press “seal” and begin the sealing process. Once you press this, you 

will be prompted on how to complete the sealing procedure. Firstly, 

place the correct plate adapter onto the shuttle, followed by the 

plate and plate weight.

This screen will show once you press the seal button, it will prompt 

you to ensure that the sealing plate is ready to be sealed, including 

the correct plate support, the corresponding plate, the correct foil, 

and finally the corresponding plate weight. Once all of these have 

been checked, you can manually push the drawer in to begin the 

sealing process. Once this is complete, the automated spring will 

return the drawer back out.

Close drawer to 

home system

Place plate foil 

and foil weight 

onto shuttle and 

close drawer

Содержание Genoseal 2060

Страница 1: ...Operating Manual 1 1300 87075 1A 2060 Genoseal Semi automatic Heat Sealer for Microplates and Tube Racks...

Страница 2: ...5 Principle of Operation 5 Installation 6 Intended Use 6 Inventory 6 Operating Environment 6 Connecting the Voltage Supply 6 Operation Procedure 7 Optimization 8 Temperature and Time Setting 8 Sealing...

Страница 3: ...s are identified and complied with Failure to do so may result in damage to the equipment and may result in personal injury In particular the user should study the contents of this guide carefully bef...

Страница 4: ...instrument If it is required to use an extension cord the cord must be earthed The Genoseal is supplied for direct connection to normal 115 230 VAC 50 60Hz supply with a variation in supply that give...

Страница 5: ...n the same way the unit was originally shipped with correct foam and crate If any damage was caused or the original crate that the Genoseal came in is lost please contact support Principle of Operatio...

Страница 6: ...at Sealing Film 2061 Plate Weight for Recessed Plates 2060 WT RC Plate Weight for Standard and Rimless Plates 2060 WT STD Plate Locators 2060 PL Accessories supplied with the Genoseal Operating Enviro...

Страница 7: ...pre cut length of film foil on the microplate and place the foil brass weight on top of the foil on the microplate Ensure that the foil is centrally placed on the microplate Note Some sealing material...

Страница 8: ...ate damage or distortion Another optimization factor to be taken into account is the surface area of the plate A plate with thin raised rims around each well will have a reduced surface area compared...

Страница 9: ...This is the main sealing screen Here you will see the main button to press seal and begin the sealing process Once you press this you will be prompted on how to complete the sealing procedure Firstly...

Страница 10: ...the settings screen Press it and adjust it using the arrows on the right side of the display Once satisfied with the set sealing time press save If you want to keep the time unchanged press cancel To...

Страница 11: ...rsists after the below actions have been carried out contact Spex for further investigation PROBLEM ACTION Genoseal will not power up after switching on Check fuse Foil will not seal properly on a mic...

Страница 12: ...As detailed under The European Low Voltage Directive 73 23 EEC amended by 93 68 EEC and the UK legislation The Electrical Equipment Safety Regulations 1994 The Declaration of Conformity is provided f...

Отзывы: