background image

Operating Manual

10

1300-87075-1A

This screen will allow you to select the settings that 

control the sealing time, temperature, ECO time and 

scan a copy of the user manual through the QR code 

displayed. After clicking settings, you will be shown this 

screen, simply click the o

 

To adjust the set temperature of the heater block and 

control the temperature that the Genoseal seals at, click 

the first option showing a thermometer. Here you can 

adjust the temperature up or down with the arrows on 

the right side of the display. Once satisfied with the set 

temperature, press “save”.

 

You can also adjust the sealing time by pressing the 

stopwatch icon on the settings screen. Press it and adjust 

it using the arrows on the right side of the display. Once 

satisfied with the set sealing time, press “save”. If you 

want to keep the time unchanged, press “cancel”.

 

To access the eco-mode timer, press the icon that shows 

the flower in a light bulb icon from the settings screen. 

This will enable you to set a time limitfor how long the 

unit is not in use before automatically turning off the 

heaters and going into eco-mode. Use this screen similar 

to how you set the temperature and time.

This will show you the QR code which is scannable with 

any QR code scannable smart phones or QR scanners 

for a desktop view. It will load a PDF in your default PDF 

viewer.

If the unit encounters an error at any stage, it will display an error 

screen prompting you to power cycle the machine or reset it. (Error 

screens may vary.)

Содержание Genoseal 2060

Страница 1: ...Operating Manual 1 1300 87075 1A 2060 Genoseal Semi automatic Heat Sealer for Microplates and Tube Racks...

Страница 2: ...5 Principle of Operation 5 Installation 6 Intended Use 6 Inventory 6 Operating Environment 6 Connecting the Voltage Supply 6 Operation Procedure 7 Optimization 8 Temperature and Time Setting 8 Sealing...

Страница 3: ...s are identified and complied with Failure to do so may result in damage to the equipment and may result in personal injury In particular the user should study the contents of this guide carefully bef...

Страница 4: ...instrument If it is required to use an extension cord the cord must be earthed The Genoseal is supplied for direct connection to normal 115 230 VAC 50 60Hz supply with a variation in supply that give...

Страница 5: ...n the same way the unit was originally shipped with correct foam and crate If any damage was caused or the original crate that the Genoseal came in is lost please contact support Principle of Operatio...

Страница 6: ...at Sealing Film 2061 Plate Weight for Recessed Plates 2060 WT RC Plate Weight for Standard and Rimless Plates 2060 WT STD Plate Locators 2060 PL Accessories supplied with the Genoseal Operating Enviro...

Страница 7: ...pre cut length of film foil on the microplate and place the foil brass weight on top of the foil on the microplate Ensure that the foil is centrally placed on the microplate Note Some sealing material...

Страница 8: ...ate damage or distortion Another optimization factor to be taken into account is the surface area of the plate A plate with thin raised rims around each well will have a reduced surface area compared...

Страница 9: ...This is the main sealing screen Here you will see the main button to press seal and begin the sealing process Once you press this you will be prompted on how to complete the sealing procedure Firstly...

Страница 10: ...the settings screen Press it and adjust it using the arrows on the right side of the display Once satisfied with the set sealing time press save If you want to keep the time unchanged press cancel To...

Страница 11: ...rsists after the below actions have been carried out contact Spex for further investigation PROBLEM ACTION Genoseal will not power up after switching on Check fuse Foil will not seal properly on a mic...

Страница 12: ...As detailed under The European Low Voltage Directive 73 23 EEC amended by 93 68 EEC and the UK legislation The Electrical Equipment Safety Regulations 1994 The Declaration of Conformity is provided f...