Operating Manual

6

1300-87075-1A

Installation

The unit should be installed on a sturdy bench to avoid any vibration during operation. When locating the unit, 

there should be enough space allowed for the drawer to move out without obstruction.

Intended Use

The Genoseal is intended to be used for the heat sealing of microplates or tubes during experimentation, 

storage, or transportation. This device is not intended to be used as a medical device.

Caution:

 Use appropriate lifting and carrying equipment available in the facility during unit movement 

from one place to another. Always use any necessary PPE during unit movement. 

Inventory

Note:

 All of the items listed below are available as spare components from Spex. 

DESCRIPTION

PART NUMBER

Plate Adapter for PCR 384-Well Format

2060-PA-384

Skirtless Plate Adapter

2060-PA-96

Heat Sealing Film

2061

Plate Weight for Recessed Plates*

2060-WT-RC

Plate Weight for Standard and Rimless Plates*

2060-WT-STD

Plate Locators*

2060-PL

* Accessories supplied with the Genoseal

Operating Environment

The Genoseal should be operated in an environment with a temperature range of 10 °C to 30 °C and a non-

condensing relative humidity range of 10% to 80%.
The unit should be kept out of drafts and air currents, as they will have an adverse effect upon the temperature 

stability of the sealing plate area. If the sealing plate area is subjected to drafts or air currents, quality or process 

time may be affected.
The unit should be placed on a secure and stable surface with a minimum of 100 mm (4-inch) clearance on all 

sides.

Connecting the Power Supply

The Genoseal utilizes a standard IEC inlet, which is fused and switched. Plug the IEC Lead in the inlet and switch 

the Rocker I.O. Switch to the On position ( | ). The display screen should illuminate.

Note: 

The system is self-regulated to sense voltage differences between 100 / 250 VAC.

Содержание Genoseal 2060

Страница 1: ...Operating Manual 1 1300 87075 1A 2060 Genoseal Semi automatic Heat Sealer for Microplates and Tube Racks...

Страница 2: ...5 Principle of Operation 5 Installation 6 Intended Use 6 Inventory 6 Operating Environment 6 Connecting the Voltage Supply 6 Operation Procedure 7 Optimization 8 Temperature and Time Setting 8 Sealing...

Страница 3: ...s are identified and complied with Failure to do so may result in damage to the equipment and may result in personal injury In particular the user should study the contents of this guide carefully bef...

Страница 4: ...instrument If it is required to use an extension cord the cord must be earthed The Genoseal is supplied for direct connection to normal 115 230 VAC 50 60Hz supply with a variation in supply that give...

Страница 5: ...n the same way the unit was originally shipped with correct foam and crate If any damage was caused or the original crate that the Genoseal came in is lost please contact support Principle of Operatio...

Страница 6: ...at Sealing Film 2061 Plate Weight for Recessed Plates 2060 WT RC Plate Weight for Standard and Rimless Plates 2060 WT STD Plate Locators 2060 PL Accessories supplied with the Genoseal Operating Enviro...

Страница 7: ...pre cut length of film foil on the microplate and place the foil brass weight on top of the foil on the microplate Ensure that the foil is centrally placed on the microplate Note Some sealing material...

Страница 8: ...ate damage or distortion Another optimization factor to be taken into account is the surface area of the plate A plate with thin raised rims around each well will have a reduced surface area compared...

Страница 9: ...This is the main sealing screen Here you will see the main button to press seal and begin the sealing process Once you press this you will be prompted on how to complete the sealing procedure Firstly...

Страница 10: ...the settings screen Press it and adjust it using the arrows on the right side of the display Once satisfied with the set sealing time press save If you want to keep the time unchanged press cancel To...

Страница 11: ...rsists after the below actions have been carried out contact Spex for further investigation PROBLEM ACTION Genoseal will not power up after switching on Check fuse Foil will not seal properly on a mic...

Страница 12: ...As detailed under The European Low Voltage Directive 73 23 EEC amended by 93 68 EEC and the UK legislation The Electrical Equipment Safety Regulations 1994 The Declaration of Conformity is provided f...

Отзывы: