In the absence of an above-mentioned overhaul, the device must be DECOMMISSIONED

, as compliance with Regulation 2017/745/EU will 

cease and, despite the CE marking, the device will no longer meet the safety requirements guaranteed by the Manufacturer at the time of 
supply. 
Spencer  Italia  srl  declines  any  responsibility  for  incorrect  operation  or  damage  caused  by  use  of  devices  that  have  not  been  regularly 
serviced. 
 
The mattress and straps shall be replaced every two years. 

Only overhaul activities carried out by specialised technicians authorised by the Manufacturer are considered valid by Spencer Italia S.r.l. 

12.4

 

SPECIAL MAINTENANCE 

Special maintenance can only be carried out by the Manufacturer, who uses internal and external technicians specialised and authorised 
by the Manufacturer itself. 

Only maintenance activities carried out by specialised technicians authorised by the Manufacturer are considered valid by Spencer Italia S.r.l.. 

The end-user can replace only the spare parts indicated in § 15. 

12.5

 

LIFE SPAN 

The device, if used as described in the following instructions, has a life span of 5 years from the date of purchase, which can be extended 
following annual overhauls. 

Overhauls  must  be  carried  out  by  the  Manufacturer,  who  uses  internal  and  external  technicians  specialised  and  authorised  by  the 
Manufacturer. 

In  the  absence  of  such  annual  overhauls,  the  device  must  be  DISPOSED  OF  ACCORDING  TO  THE  INFORMATION  IN 

PARAGRAPH 16 AND NOTIFICATION MUST BE GIVEN TO THE MANUFACTURER.  

 

The life span may be extended at the sole discretion of the Manufacturer or authorised centre if the safety requirements of the device are 
still met. 

Spencer  Italia  S.r.l.  will  accept  no  responsibility  for  incorrect  operation  or  damage  caused  by  the  use  of  devices  that  have  not  been 
overhauled by the Manufacturer or authorized centre, or that have exceeded the maximum allowable life span. 

13.

 

TROUBLESHOOTING TABLE 

 

PROBLEM 

CAUSE 

REMEDY 

The leg unlocking levers don’t work properly or 

are difficult to operate. 

The moving mechanisms are 
damaged. 

Immediately  remove  the  device  from  service  and 
contact the service centre. 

Means of connection between 
components have been lost. 

Immediately  remove  the  device  from  service  and 
contact the service centre. 

The stretcher is not correctly attached to the 
fixing system. 

Wear or damage of the  
components that constitute the 
locking mechanisms. 

Immediately  remove  the  device  from  service  and 
contact the service centre. 

The fifth wheel has not been 
properly inserted in the fixing 
system. 

Position  the  stretcher  properly  and  check  that  the 
fifth  wheel  enters  the  dedicated  part  on  the  fixing 
system. 

Damage to the structure 

Improper use 

Immediately  remove  the  device  from  service  and 
contact the service centre. 

It is not possible to set the stretcher in 
intermediate position. 

The moving mechanisms are 
damaged. 

Immediately  remove  the  device  from  service  and 
contact the service centre. 

Something is interfering with the 
moving systems. 

Make  sure  that  nothing  is  interfering  with  the 
mechanisms. 

Levers have not been properly 
operated. 

Carefully  follow  the  instructions  to  place  the 
stretcher in the intermediate position. 

The front legs do not lock when unloading from 
the medical vehicle. 

The moving mechanisms are 
damaged. 

Immediately  remove  the  device  from  service  and 
contact the service centre. 

The height of the loading floor is 
not suited to the device, the safety 
height is not being respected. 

Adjust  the  loading  floor  height  to  meet  the 
requirements  described  in  this  manual.  If  the 
loading  floor  does  not  allow  adjustments, 
immediately  remove  the  device  from  service  and 
contact the service centre. 

When the stretcher release lever is operated, the 
stretcher does not move from the fixing system 
and   
remains attached. 

The magnet controlling the release, 
located inside the fifth wheel 
holder, has moved or been lost. 

Manually  unlock  the  stretcher  and  finish  the 
unloading  operation.  At  the  end  of  service,  check 
the  positioning  of  the  magnet,  then  restore  its 
original  position.  If  the  problem  persists, 
immediately  remove  the  device  from  service  and 
contact the service centre. 

Содержание Carrera Tec

Страница 1: ... ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy SPENCER ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy www spencer it support spencer it e mail info spencer it service service spencer it Ph Tel 39 0521 541111 Quality System EN ISO 9001 EN ISO 13485 ...

Страница 2: ...2 P1 P2 P3 P4 ...

Страница 3: ...3 P3 P4 P5 P6 P7 P8 P9 P10 ...

Страница 4: ...4 P12 P13 P14 P15 P16 SHARP SHARP SL P17 P18 P19 P20 P21 P22 P23 ...

Страница 5: ...5 P24 P25 P26 P27 ...

Страница 6: ... P14 and P15 11 11 5 Trendelenburg Board Image P17 11 11 6 Trendelenburg Fowler Board Image P18 11 11 7 Backrest adjustment Image P19 12 11 8 Changing the stretcher length ACTIV only Image P20 P21 12 11 9 Lowering the stretcher Image P22 P23 12 11 10 Telescopic handle operation Image P26 13 11 11 Lifting the stretcher with the patient 13 11 12 Spine board housing Carrera Pro only Image 27 13 12 CL...

Страница 7: ...Rev 0 04 05 2021 Code CCI5297 Copyright Spencer Italia S r l All rights reserved No part of the document may be photocopied reproduced or translated into another language without prior written consent from Spencer Italia S r l ...

Страница 8: ...instructions are the continuous practice of use under the supervision of competent and trained personnel Regardless of your level of experience with similar devices in the past you should carefully read and understand the contents of this manual before installing operating or servicing this product In case of any questions please contact Spencer Italia S r l for the necessary clarifications The pr...

Страница 9: ...y will be void as the device can no longer be traced If the assigned Lot SN cannot be traced the device must be reconditioned provided only under the responsibility of the manufacturer Regulation 2017 745 EU requires manufacturers and distributors of medical devices to keep track of their location If the device is in a location other than the address to which it was shipped or sold or if it was do...

Страница 10: ...ht which weighs on each operator must comply with local health and safety requirements Avoid contact with sharp objects Device installation must be carried out by qualified personnel trained and enabled by Spencer Italia S r l The times and methods of such shall be agreed upon between the customer and our Sales Offices Operating temperature from 10 C to 50 C Storage The product must not be exposed...

Страница 11: ...e utmost attention to any obstacles water ice debris etc present on the route as they may cause the operator to lose balance and compromise proper functioning of the device If you cannot clear the way choose an alternative route Condensation water ice and accumulations of dust can affect correct operation of the device making it unpredictable and causing a sudden change in the weight the operators...

Страница 12: ...sual circumstances the presence of multiple operators is recommended not just 2 as expected in standard conditions The ability of each operator must be assessed before the definition of rescuer roles in use of the stretcher 7 RESIDUAL RISK The residual risks listed below have been identified solely with reference to the intended use of the device Use by untrained personnel may result in injury to ...

Страница 13: ... fixing without intervening on the fixing itself Nylon PE Steel 8 Front leg When closed it allows the stretcher to be loaded into the ambulance two wheels are attached to the stretcher which are necessary for its movement Steel 23 Sharp system drive handle if present Control part of the Sharp System Nylon Steel 9 Front wheel needed to move the stretcher PU 24 TEC TH Rear leg unlocking lever Nylon ...

Страница 14: ...operation and damage to the patient or rescuers and will void the warranty and release the Manufacturer from all liability Carrera stretchers are available in various heights to be assessed before purchase in relation to the loading configuration on the vehicle NOTE Please contact the manufacturer for installations requiring the use of accessories The type of stretcher should be chosen taking into...

Страница 15: ...ng lever Unlock sidebar opening when right and left are pressed at the same time G Loading trolley unlocking sleeve Only available on ACTIV versions unlocks movement of the loading trolley when activated allowing the stretcher to be shortened H Back lift piston knob Pulling it releases the stop to allow the backrest to be returned to a lower position than it is currently in I Telescopic handle unl...

Страница 16: ...AL STRETCHER MOVEMENTS NEVER ENGAGE THE LEG LOCKING LEVERS AS DOING SO COULD CAUSE THE STRETCHER TO FALL TO THE GROUND THESE LEVERS MUST ONLY BE ENGAGED TO LOAD THE STRETCHER ON THE AMBULANCE OR TO REACH THE INTERMEDIATE HEIGHT AS DESCRIBED ABOVE To unload the stretcher from the medical vehicle If the stretcher is used with the Sensor Lock fixing system release the stretcher by engaging the rear l...

Страница 17: ...d patients onto the stretcher at intermediate height Engage the rear leg unlocking lever hold it in place and lift the stretcher slightly to unlock the rear leg moving mechanism Lower the stretcher about 10 cm release the unlocking lever and support the stretcher until the intermediate height position is reached Make sure that the stretcher has reached a stable position Engage the brakes on the re...

Страница 18: ...injury to the patient or the operator by applying excessive force on the device 12 CLEANING AND MAINTENANCE Spencer Italia S r l declines all responsibility for any direct or indirect damage which is the consequence of improper use of the product and spare parts and or in any case of any repair carried out by a person other than the Manufacturer who uses internal and external technicians authorise...

Страница 19: ...ce maintenance must be carried out by operators with specific qualifications training and instruction in the use and maintenance of the device The operator must wear suitable personal protective equipment such as gloves goggles etc during all checking maintenance and cleaning operations Checks to be carried out before and after each use and at deadline indicated above are as follows General functi...

Страница 20: ...g levers don t work properly or are difficult to operate The moving mechanisms are damaged Immediately remove the device from service and contact the service centre Means of connection between components have been lost Immediately remove the device from service and contact the service centre The stretcher is not correctly attached to the fixing system Wear or damage of the components that constitu...

Страница 21: ...faulty Check the fixing system power supply If the problem persists immediately remove the device from service and contact the service centre The stretcher is used with a fastener that does not provide automatic release If necessary request a different fixing system ...

Страница 22: ...ow oxygen tank for stretcher siderail fixing CB09026C Siderail fixing support for tanker oxygen tank ST00497B DNA Strap chest strap with re winding system ST00498B DNA Strap with re winding system ST00499B STX 499 UNIVERS ADJUST 4 PC CHEST STRAP ST00592A STX 592 YELLOW METAL 2 PC METAL HOOK STRAP 15 SPARE PARTS ST70002A STX 702 Two piece metal black reflex strap black ST70000A QMX 777 Anatomical s...

Отзывы: