XLG/XLF MANUAL
12
118020-001 REV E
1
JB1
M
PS Common
It is recommended that analog signals be isolated
5
Current
Monitor
2
3
4
M
Monitor
Voltage
6
7
Z out = 4.99k ohms ±0.1%
Z out = 4.99k ohms ±0.1%
via isolation amplifiers.
All cables should be shielded with the shields
being returned to the chassis ground of the
High Voltage Power Supply.
Figure 3.4 Remote Monitor Test Point Designations
It is recommended to use relay contacts for S1 and S2.
Relays should be located as close as possible to the high
voltage power supply. Coils should be driven form
isolated sources. Signals are at ¢ 15Vdc, 25mA max, and
are only to be used for contact closure.
S1 must be closed to enable HIGH VOLTAGE ON.
S1
S2
S1=Remote High
Voltage Off
High Voltage Power Supply.
to the chassis ground of the
14
JB1
Remove jumper for remote
15
16
17
Voltage On
S2=Remote High
high voltage control
All cables should be shielded
with the shields being returned
Figure 3.5 Remote Control of High Voltage ON and High Voltage OFF Interface