DF3/FF3 MANUAL 

118018-001 

Rev E

 

 

 

Line Regulation:  0.005% of rated 
output over specified input range. 

 

Temperature Coefficient:

  50 ppm/˚C (20 

ppm/˚C optional). 
 

Long term Stability:  

0.01% / 8 hours. 

 

Current Regulation:

 

 
 

Load regulation:  0.01% of rated 
output for a 10-60kV change. 

 
 

Line regulation:  0.005% of rated 
output over specified input range. 

 

Temperature Coefficient: 

 50 ppm/C 

 

Long term Stability:  

0.01% / 8 hours. 

 

Ripple:  

0.03% below 1kHz, 0.75% rms above 

1kHz. 
 

Filament Voltage:  

14Vac (dc filament 

optional). 
 

Filament Current:  

12A ac. 

 

Size:  

5 ¼” (13.26cm)H x 19” (48.26cm)W x 

22” (55.88cm)D. 
 

1.4

 

STANDARD FEATURES 

The DF3/FF3 series incorporates extensive 
standard features designed to optimize user 
control and safety. 
 

KV AND MA/FILAMENT CURRENT 
RAMP CIRCUITS: 

This feature provides for a 

gradual rise for kV, mA, and filament current. 
This feature is designed to limit voltage shock 
and filament shock to the X-ray tube. The kV 
ramp rate is approximately 10 seconds. The 
filament current is slowly increased until the 
desired mA level is achieved. This time is 
typically 20 seconds for full mA output. These 
ramp conditions are started at the initial X-RAY 
ON control signal or when the kV control, 
(front panel potentiometer or remote 
programming), drops below the kV MIN level. 
The kV MIN level is typically set at 8kV. The 
kV MIN level is adjustable by an internal 
resistor valve. Other kV MIN levels can be 
accommodated. Consult Spellman Engineering 
for further information. 

FAIL-SAFE, DUAL INDICATOR LAMPS:

 

The X-ray on lamp utilizes a sensing circuitry for 
detecting proper operation of the lamp for the 
indicator. If the X-ray on lamp is not functioning, 
the power supply will not generate high voltage. If 
the X-ray on lamp is depressed and the indicators 
do not illuminate, the lamps are defective and 
should be replaced in order to generate high 
voltage. Control power and X-ray off lamps are 
also dual type lamps for added reliability. 

 

 

INTERNAL FAULT PROTECTION AND 
DIAGNOSTICS:

 The DF3/FF3 series 

continually monitors internal circuits critical to 
the proper operation of the power supply and X-
ray tube. The diagnostic circuits utilize 
sophisticated Programmable Logic Devices 
(PLD’s).  The diagnostic programming is a 
software based, decision-making technology. 
Flexible software provides very sophisticated 
decision-making capabilities within the power 
supply.  Also, a variety of user configurations can 
be achieved by custom software written by 
Spellman Engineering.  If a specific fault 
condition or diagnostic is required, consult 
Spellman High Voltage Engineering Department. 

1.4.1 REMOTE 

OPERATING 

FEATURES 

REMOTE PROGRAMMING:

 Allows remote 

adjustment of the output voltage, output current, 
filament current limit, and maximum tube power 
via an external voltage source. 
 

REMOTE MONITOR:

 Allows remote 

monitoring of the output voltage and output 
current. 
 

REMOTE X-RAY ON/OFF CONTROL:

 

Allows remote control of X-ray ON and X-ray 
OFF functions. Signals are also provided for 
remote indication of X-ray ON status. This can be 
used for a variety of water control or remote lamp 
functions. 
 

REMOTE POWER ON:

 Allows remote control 

of Control Power ON/OFF function. This feature 
also allows remote reset of the power supply in 
case of a fault shutdown. 
 

EXTERNAL INTERLOCK:

 Interlock 

connections are provided for a variety of water 
flow and safety functions. The unit will not 
operate unless the interlock is closed. Opening the 
interlock circuit will cause the unit to shut off. 
 

Содержание DF3 Series

Страница 1: ...SERIAL DATE SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION 475 Wireless Blvd Hauppauge New York 11788 1 631 630 3000 FAX 1 631 435 1620 E mail sales spellmanhv com Website www spellmanhv com DF3 FF3 MANUAL 118019 001 Rev C DF3 FF3 SERIES ...

Страница 2: ... aware of the electrical hazards WARNING note in the text call attention to hazards in operation of these units that could lead to possible injury or death CAUTION notes in the text indicate procedures to be followed to avoid possible damage to equipment Copyright 2000 Spellman High Voltage Electronics Corporation All Rights Reserved This information contained in this publication is derived in par...

Страница 3: ...spannungsnetzteiles benötigen ca 5 Minuten um sich zu entladen Erden Sie sich nicht und arbeiten Sie nicht in feuchter oder nasser Umgebung Notwendige Reparaturen können es erforderlich machen den Gehäusedeckel während des Betriebes zu entfernen Reparaturen dürfen nur von qualifiziertem eingewiesenem Personal ausgeführt werden WARNING im folgenden Text weist auf gefährliche Operationen hin die zu ...

Страница 4: ...de l alimentation Ne vous mettez pas à la masse ou ne travaillez pas sous conditions mouillées ou humides La maintenance peut nécessiter l enlèvement du couvercle lorsque l alimentation est encore allumée Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques Dans le manuel les notes marquées WARNING attire l attention sur les risques lors de la m...

Страница 5: ...carica della capacità interna dell alimentatore ad alta tensione Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d umidità SICUREZZA NELLA MANUTENZIONE Manutenzione potrebbe essere richiesta rimuovendo la copertura con apparecchio acceso La manutenzione deve essere svolta da personale qualificato coscio dei rischi elettrici Attenzione alle AVVERTENZE contenute nel ...

Страница 6: ...SSEMBLY 20 4 8 REAR PANEL INTERFACE PWB 20 4 9 FILAMENT POWER PWB 21 4 10 OPTIONS 21 5 OPTIONS 22 5 1 208V AC THREE PHASE INPUT 22 5 2 HIGH STABILITY 22 5 3 220VAC FAIL SAFE CIRCUIT 22 5 4 12VDC FAIL SAFE CIRCUIT 22 5 5 24VDC WATER FLOW SIGNAL 22 5 6 220VAC WATER FLOW SIGNAL 22 5 7 DUAL FILAMENT SUPPLY 22 5 8 4KV OUTPUT 22 5 9 ELAPSED TIME METER 22 5 10 LARGE COMMON 22 5 11 NON STANDARD SLOW START...

Страница 7: ...AGRAM FOR JB1 LOCAL CONTROL 10 FIGURE 3 2 REMOTE CONTROL 12 TABLE 3 1 LOCAL REMOTE DIP SWITCH SETTINGS 12 FIGURE 3 3 REMOTE MONITOR 13 FIGURE 3 4 REMOTE ON OFF CONTROL 14 FIGURE 3 5 REMOTE STATUS INDICATOR 15 FIGURE 3 6A REMOTE FAIL SAFE INDICATORS 24V DC 16 FIGURE 3 6B REMOTE FAIL SAFE INDICATORS 220V DC 17 FIGURE 3 7D REMOTE WATER SELENOID CONTROL 17 TABLE 4 1 DF3 FF3 ASSEMBLIES 19 APPENDIX SPEC...

Страница 8: ...ity and the filament circuits are referenced to cathode ground potential grounded filament The X ray tube voltage tube emission current tube filament current limit and maximum tube power are all continuously adjustable front panel or remote controls The power supplies operate from a variety of line voltages and are fan cooled requiring no water cooling Custom designed units for single use or OEM a...

Страница 9: ...ning the power supply will not generate high voltage If the X ray on lamp is depressed and the indicators do not illuminate the lamps are defective and should be replaced in order to generate high voltage Control power and X ray off lamps are also dual type lamps for added reliability INTERNAL FAULT PROTECTION AND DIAGNOSTICS The DF3 FF3 series continually monitors internal circuits critical to th...

Страница 10: ... OVER VOLTAGE Indicates the over voltage protection circuitry has caused the power supply to turn off Over voltage is internally set to 105 110 of rated output voltage OVER CURRENT Indicates the output current has exceeded 110 of rated output current The over current level is internally set OVER POWER indicates the output power has exceeded the front panel or remote programmed power trip level ARC...

Страница 11: ...g at 220Vac replaces WFS 220V standard circuit FSI 12VDC Remote Fail Safe circuit operating at 12Vdc 24Vdc standard WFS 24V PO 24Vdc water flow signal switched with power ON switched with X Ray ON is standard WFS 220V PO 220V ac water flow signal switch with power ON switched with X Ray ON standard Not available with FSI 220V DFS Dual Filament supply FF3 Only 4kW 4kW maximum output power ETM Elaps...

Страница 12: ...chassis for visible damage Fill out and mail the Warranty Registration card accompanying the unit Standard Spellman DF3 FF3 high voltage power supplies and components are covered by warranty Custom and special order models with an X suffix in the model number are also covered by warranty 2 2 MECHANICAL INSTALLATION Units in the DF3 FF3 series have front panel holes for standard EIA rack mounting T...

Страница 13: ...DF3 FF3 MANUAL 6 118018 001 Rev E Figure 2 1 DF3 FF3 DIMENSIONS ...

Страница 14: ...d with a different input voltage C Proper Grounding Technique The chassis of the high voltage power supply must be grounded to the local earth ground and also to the tube housing ground See Figure 3 1A 3 1B and 3 1C for typical operating setup A three wire cable is provided for connecting to the AC supply The Green wire is for chassis safety ground connection D Connect the X ray cable to the X ray...

Страница 15: ...DF3 FF3 MANUAL 8 118018 001 Rev E FIG 3 1A DF3 FF3 FRONT PANEL LAYOUT ...

Страница 16: ...DF3 FF3 MANUAL 9 118018 001 Rev E FIG 3 1B DF3 TYPICAL OPERATING SETUP FIG 3 1C FF3 TYPICAL OPERATING SETUP ...

Страница 17: ... power supply Relay coil should be driven from an isolated source All cables should be shielded with shields being returned to chasses ground E1 of the High Voltage power supply 1 11 20 5 2 3 22 23 10 S1 External Lock Failsafe Interlock 28V 1W Lamp is energized when the power on switch 24V 10mA is provided JB1 FIGURE 3 1D WIRING DIAGRAM FOR JB1 WHEN NO REMOTE OPERATION IS DESIRED ...

Страница 18: ...nal ramp circuits Once kV and mA levels are reached 2 20 seconds adjust kV and mA for desired output L To terminate the generation of X rays depress the X ray OFF switch M To turn off the power supply depress the Control Power ON Switch out WARNING After turn off do not unplug the output cable until the internal capacitance is discharged approximately 1 minute WARNING The voltmeter on the power su...

Страница 19: ...t to 10 volts full rated output the desired output can be selected See Figure 3 2 for wiring diagram and specifications for remote program interface requirements V Pwr Limit Com OL Com Pwr Limit OL Ref 0 10Vdc V Filament Current Limit Com Filament Current Limit 0 10Vdc 3 Vin must be referenced to chassis common potential or within 1V of ground potential 2 Zin 10Meg 1 Input Instrumentation amplifie...

Страница 20: ... always positive regardless of the output polarity where 0 to 10V 0 100 of output Accuracy is within 0 5 All test points have output impedance at 1Kohm See Figure 3 3 for remote monitor interface 38 47 JB1 M KV Monitor Mon Common 2 All signals 0 10Vdc 0 100 of specified maximum outputs 1 Metering Circuit should be high Z type 10 Meg for accurate measurement NOTES Customer Interface High Voltage Po...

Страница 21: ...erenced to chassis ground 2 Relay coil Z 178 1 The power supply provides 5Vdc for coil voltage however for full isolation 4 Relay contacts or switches can be used instead of transistors NOTES Customer Interface High Voltage Power Supply 5 X Ray OFF Transistor momentarily for constant X Ray OFF OFF for X Ray OFF or 4 X Ray ON 3 INTLK Closed for INTLK 11 8 RESET ON RESET Momentary transistor Optiona...

Страница 22: ...iode of an opto coupler See Figure 3 5 for recommended interface circuits JB1 2 Low level on transistor status active 1 100 mA max collector current on internal power supply transistors NOTES Customer Interface High Voltage Power Supply 12 1 4W 5Vdc X Ray ON 1 4W 19 Power Supply Fault 5Vdc 1 4W 16 Filament Current Limit 5Vdc 1 4W 13 Overvoltage 5Vdc 1 4W 15 Overpower 5Vdc 1 4W 17 mA Current Limit ...

Страница 23: ...C signals are provided These signals are switched ON when X RAY ON is switched on These signals can also be controlled independently when the power supply is switched ON This is accomplished on the rear panel INTERFACE PWB JP2 controls the 24V signal and JP3 controls the 220V signal See Figure 3 7 for remote water flow connections External interlock connections can also be used for water flow inte...

Страница 24: ...INDICATOR 220v AC High Voltage Power Supply Customer interface Switched 220VAC Solenoid 220VAC 2 0Amp max IEC 320 Connector F1 2 5A 2 5A F2 Notes 1 F1 and F2 are located on the Filament Interface PWB 2 The 220VAC signal switch to the IEC 320 connector is normally controlled by X RAY ON switch The signal can be controlled by the power on switch by JP2 connector on the Rear Panel Interface PWB Consu...

Страница 25: ... recommended practices for safety in high voltage testing 510 1983 4 2 LINE RECTIFICATION AND FILTERING Basic AC line rectification provides the DC voltage for the high frequency inverter and the filament supply The FILTER CAP ASSEMBLY and main chassis components comprise the circuits for line voltage rectification The line input voltage can vary from 180 up to 242 volts within the DF3 FF3 series ...

Страница 26: ...The IGBT PWB provides a variety of control and diagnostic functions R13 R14 are the charging resistors previously described These resistors limit the initial charging current for the capacitors used for lone rectification U1 and its associated circuitry monitor the voltage across the charging resistor If any abnormal conditions appear circuitry will shutdown the power supply R38 is a bleeder resis...

Страница 27: ...rcuits of U8 Arc sense circuits T1 and U6 provide sensing quench and indication of arc conditions Current feedback from the high voltage rectifier is sent to sense resistors R56 R24 and R57 U10 provides gain adjustment MAFDBK CAL An over voltage protection divider located in the HIGH VOLTAGE ASSEMBLY provides protection in the unlikely event of a main control loop failure U14 provides gain adjustm...

Страница 28: ...e filament isolation transformer The filament isolation transformer secondary is connected to the output connector For the DC filament option the transformer secondary is connected to a full wave rectifier and then to the output connector 4 10OPTIONS Due to many variations of models and options provided in the DF3 FF3 series details of actual circuits used may differ slightly from above descriptio...

Страница 29: ...PPLY DFS A dual filament supply is available on FF3 models only This option allows the FF3 to power Varian DTS 68 X Ray tubes and equivalents 5 8 4kW OUTPUT 4kW 4kW maximum power is available On FF3 units maximum current is limited to 80 mA 100mA is standard on 3kW FF3 5 9 ELAPSED TIME METER ETM An elapsed time meter is provided It displays cumulative hours of X RAY ON 5 10LARGE COMMON LC The stan...

Страница 30: ...EROUS Only qualified personnel should perform these tests High voltage test procedures are described in Spellman Bulletin STP 783 Standard Test Procedures for High Voltage Power Supplies Copies can be obtained from the Spellman Customer Service Department Test equipment including an oscilloscope a high impedance voltmeter and a high voltage divider such as the Spellman HVD 100 or HVD 200 is needed...

Страница 31: ... serial number of the power supply 7 2 CORRESPONDENCE AND ORDERING SPARE PARTS Each Spellman power supply has an identification label on the rear of the chassis that bears its model and serial number When requesting engineering or applications information please state the model and serial number of the power supply If specific components or circuit sections are involved in the inquiry it is helpfu...

Страница 32: ...he customer The customer shall pay for shipping to and from Spellman A preliminary estimate for repairs will be given by phone by Customer Service A purchase order for this amount is requested upon issuance of the RMA Number A more detailed estimate will be made when the power supply is received at Spellman Repair Center In the event that repair work is extensive Spellman will call to seek additio...

Страница 33: ...To obtain information on Spellman s product warranty please visit our website at http www spellmanhv com en About Warranty aspx ...

Отзывы: