118
M2i.30xx / M2i.30xx-exp Manual
Trigger
Mode Gated Sampling
Channel window trigger for leaving signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow.
When the signal leaves the window from the inside to the
outside, the gate will start.
When the signal enters the window from the outside to the
inside, the gate will stop.
As this mode is purely edge-triggered, the signal within the
window at the cards start time does not trigger the board.
Channel window trigger for inner signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow.
When the signal enters the window from the outside to the
inside, the gate will start.
When the signal leaves the window from the inside to the
outside, the gate will stop.
As this mode is level-triggered, the signal inside the window
at the cards start time does trigger the board.
Channel window trigger for outer signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow.
When the signal leaves the window from the inside to the
outside, the gate will start.
When the signal enters the window from the outside to the
inside, the gate will stop.
As this mode is level-triggered, the signal outside the win-
dow at the cards start time does trigger the board.
Register
Value
Direction
set to
Value
SPC_TRIG_CH0_MODE
40610
read/write
SPC_TM_WINLEAVE
00000040h
SPC_TRIG_CH0_LEVEL0
42200
read/write
Set it to the upper trigger level relatively to the channel’s input range.
board dependant
SPC_TRIG_CH0_LEVEL1
42300
read/write
Set it to the lower trigger level relatively to the channel’s input range.
board dependant
Register
Value
Direction
set to
Value
SPC_TRIG_CH0_MODE
40610
read/write
SPC_TM_INWIN
00000080h
SPC_TRIG_CH0_LEVEL0
42200
read/write
Set it to the upper trigger level relatively to the channel’s input range.
board dependant
SPC_TRIG_CH0_LEVEL1
42300
read/write
Set it to the lower trigger level relatively to the channel’s input range.
board dependant
Register
Value
Direction
set to
Value
SPC_TRIG_CH0_MODE
40610
read/write
SPC_TM_OUTSIDEWIN
00000100h
SPC_TRIG_CH0_LEVEL0
42200
read/write
Set it to the upper trigger level relatively to the channel’s input range.
board dependant
SPC_TRIG_CH0_LEVEL1
42300
read/write
Set it to the lower trigger level relatively to the channel’s input range.