clicking the icon on the SpecWare toolbar, or from the WatchDog 

Manager screen.

 

 

Select the logging interval (ranging from 1 to 120 minutes), and 

ensure  all  channels  are  enabled  before  clicking 

Ok

” 

to  launch 

the logger.

 

 

IMPORTANT!    The  A160  does  not  begin  logging  until  you 

press and hold the 

LOG

” 

button for two seconds. 

The LCD 

will  alternate  between 

rEC

” 

and  the  CO2  value  to  indicate  it  is 

logging.

 

 

Downloading The Loggers

 

The download sequence is initiated by selecting the 

Get Watch-

Dog  A

-

Series  data

” 

option  from  the  Logger  menu,  clicking  the 

icon  on  the  SpecWare  toolbar,  or  from  the  WatchDog  Manager 

screen.. When the download is complete, the data will be saved 

in the location defined for that logger. 

End

” 

will flash on the LCD 

to show the download was complete. Press and hold the 

ESC

” 

button to return to the normal display. 

 

 

Charts/Reports

 

Whenever a file is opened, SpecWare will automatically generate 

a chart. You will be given the choice of which parameters are to 

be  plotted.  The  Tools  menu  gives  you  access  to  a  variety  of 

standard and user

-

defined (SpecWare Pro only) reports.

 

Содержание WatchDog 3611A

Страница 1: ...A160 Temp RH CO2 Data Logger PRODUCT MANUAL Item 3611A...

Страница 2: ...SpecWare Software 8 Troubleshooting 10 Logger Capacity 15 999 data points 5 333 per sensor CO2 Range 0 to 9999 ppm CO2 Accuracy 50ppm 5 of reading from 0 to 2000 ppm else unspecified Pressure Depende...

Страница 3: ...ger and keep it working accurately for many years Read this manual thoroughly in order to get effective performance from your data logger The A160 logger also has a comprehensive LCD display with cont...

Страница 4: ...meter measures continuously and updates the LCD every second With a significant environmental change the CO2 sensor requires 2 minutes and the RH sensor 10 minutes to respond Note Do not hold the mete...

Страница 5: ...values or ESC to exit without saving Press SET again to enter P1 3 to set the CO2 alarm lower limit exceeding this triggers the alarm Typically the upper level for NORMAL and the lower level for the...

Страница 6: ...software see p 9 What ever value is shown on the LCD will be over ridden when the log ger is launched CO2 Manual Calibration Manual calibration should be done outdoors on a sunny day with fresh air an...

Страница 7: ...river prior to its use The driver is included on the SpecWare Software installation CD for SpecWare 9 Pro or Basic Version 9 01 Build 206 or greater as well as on the SpecWare 9 update page at www spe...

Страница 8: ...is read by SpecWare an initialization screen will open Enter the serial number printed on the bottom of the logger Because you have an A160 the A125 A130 options are ignored Logger Location Names Spe...

Страница 9: ...ce is initiated by selecting the Get Watch Dog A Series data option from the Logger menu clicking the icon on the SpecWare toolbar or from the WatchDog Manager screen When the download is complete the...

Страница 10: ...the AC adapter Shake any water out of the A160 Set the A160 in a ventilated area for at least a day preferably more Do not plug it in until you are confident that any water has evaporated Water damage...

Страница 11: ...appears then con tact your distributor regarding repair E03 Air temperature is above the upper limit Put the meter in regular room temperature for 30 minutes If this still appears then con tact your d...

Страница 12: ...CISPR11 EN 61000 4 2 1995 A1 1998 A2 2001 EN 61000 4 3 2006 This product is warranted to be free from defects in material or work manship for one year from the date of purchase During the warranty pe...

Отзывы: