Troubleshooting
The A160 logger will not power on.
-
Check that the AC adapter is plugged in to an outlet and the
A160.
-
Use a paper clip or toothpick to press
“
reset
”
on the base of the
logger.
The readings on the LCD don
’
t change.
Check to see if the display is in maximum or minimum mode.
“
Bat
”
and the green LED keep flashing.
The adaptor output voltage is not 5VDC (±10%), 500mA. Please
ensure you are using the correct adapter.
Why won
’
t the logger connect with the computer?
The COM Port settings may need to be changed in SpecWare
software.
I accidentally sprayed water on my A160.
If you haven
’
t done so already, unplug the AC adapter. Shake
any water out of the A160. Set the A160 in a ventilated area for
at least a day, preferably more. Do not plug it in until you are
confident that any water has evaporated. Water damage is not
covered under warranty.
LCD Diagnostic Codes
Code
Problem
Solution
CO
2
Reading
E01
The CO
2
sensor is
damaged.
Contact your distributor regarding repair.
E02
The CO
2
reading is
under the lower limit.
Recalibrate the A160. If this reappears, then
contact your distributor regarding repair.
E03
The CO
2
reading is
above the upper limit.
Put the A160 in fresh air for 5 minutes. If it
still appears, recalibrate the unit. If this fails,
contact your distributor for repair.
E17
ABC mode (auto
-
calibrate) of CO
2
sen-
sor has failed and
might cause incorrect
CO
2
readings.
Perform a manual calibration as per manual
page 6.