Electrical Specifications
TM 787100 Rev (Af), May 2019
35
5
Technical Specifications
5.1
Electrical Specifications
•
Operating Voltage
: 18
–
32VDC
•
Power Consumption
:
Max. 25mA in standby
Max. 50mA in alarm
•
Electric Input Protection
:
•
The input circuit is protected against voltage-reversed polarity,
voltage transients, surges, and spikes according to
MIL-STD-1275A.
Electrical Outputs
•
Dry Contact Relays:
Table 15: Contact Ratings
Relay
Name
Type
Normal
position
Maximum Ratings
Alarm
SPST
N.O.
2A at 30VDC
Fault *
SPST
N.C.
2A at 30VDC
* The fault relay is normally energized and the contact is closed during
normal operation of the detector. The contact is open at fault condition or
low voltage.
•
4
–
20mA current output:
The 4
–
20mA is an isolated sink option.
The maximum permitted load resistor is 600 Ohm.
Table 16: 0
–
20mA Current Output
STATE
Output
Fault
0 + 0.9mA
Bit Fault
2mA±10%
Normal
5mA±10%
Warning
10mA±10%
Alarm
15mA±10%