To check the PkDf and PkD1 values:
For Channel A, press the
#
key followed by the
4
key to enter
Mode 4
(System
Diagnostic Parameters - Channel A) or press the
#
key followed by the
5
key
to enter
Mode 5
(System Diagnostic Parameters - Channel B). A brief wait
screen will appear as the diagnostic data for Channel B is polled.
18.
Compare the values for
PkDf
and
PkD1.
If the difference is more
than 4, reset the peak tracking function.
To reset the Peak Tracking Function:
1.
Press the
#
key followed by the
2
key (Channel A) or the
#
key
followed by the
3
key (Channel B). The LCD prompts for a numeric
password. Enter the user password (
3142
) on the keypad, then
PP2F:8190 PkDf:34.00
Powr:2538 PkD1:34.00
Indx: 301
Zero: -24
<Diagnostic for CHB>
Please wait...
PP2F:8190 PkDf:47.00
Powr:2028 PkD1:47.00
Indx: 253 T:
50.7
Zero: -16 P:
955.7
<SET PARAMETER MODE>
Enter password
v2.40 08-12-08 DM
UK Distibution by IMA Ltd