Modbus Address
The
Modbus Address
parameter sets the analyzer address for when the
analyzer is used as a Modbus device. Addresses from 1 to 250 can be used.
Modbus Mode
The
Modbus Mode
parameter sets the communications protocol for the RS232
port. There are three choices:
0
for turning the Modbus capabilities off and
defaulting to generic serial output as described in
“Receiving Serial Data
(RS-232 Output)”
on page 4-1;
1
for enabling the analyzer to respond to
Gould Modbus RTU function codes 3, 6 and 16; and
2
for enabling the analyzer
to respond to Daniel Modbus RTU function codes 3, 6 and 16.
User Analyzer ID
The
User Analyzer ID
parameter assigns a numeric identification to the data
enabling SCADA or DCS systems to associate data with a particular analyzer.
<SET PARAMETER MODE>
Modbus Address
1
Enter a value(1-250)
<SET PARAMETER MODE>
Modbus Mode
0
0:Off 1:GMR 2:DMR
<SET PARAMETER MODE>
User Analyzer ID
0
Enter a value