SYSTEM INSTALLATION & INITIALIZATION:
1. Most of the components of this system are plastic and are subject to
damage by ultraviolet light from the sun and other sources such as
metal halide lighting.
2. Avoid installing this unit in an area where it may be subjected to bright
light or direct sunlight, as algae is more likely to thrive inside the clear
filter housings when exposed to bright light. The unit must be kept
out of areas that are subject to freezing temperatures.
3. High temperatures greater than 113° F (45° C) must be avoided. If the
unit is used outside, avoid putting the system in direct sunlight or
connecting it to a garden hose that may be exposed to sunlight.
4. Attach the black tap water tubing to the left side of the Sediment Filter
Housing (See page 6 or 7).
5. Attach the yellow drain line to the “Y”-connector at the rear of the system.
This “Y”-connector already has a yellow and red tube attached to it.
6. Attach the blue product line to the open port of the ASO Valve (see
page 8).
7. Attach the black tubing and elbow(s) to the membrane housing
cap(s).
8. Attach the garden hose adapter
(connected to the black tubing)
to
your cold water source. Never run hot water (greater than 113° F /45° C)
through the system.
9. A provided ball valve can be located at a convenient place on the
black tap water line. Cut the black tubing in two and reconnect the
two cut ends with the ball valve.
10. Place the
yellow concentrate (waste) tubing
and the
blue purified
(product) water tubing
temporarily into a drain. Do not restrict flow
from these lines.
11. Open the cold water supply valve. The pressure should not exceed 80 psi .
12.
Check the system to ensure that all fittings are tight and leak-free
before leaving the system unattended. (If anything is leaking,
contact SpectraPure for assistance.)