QUERY
This command displays the monochromators status.
To CM110/112:
<56>
D
<Query Byte>
From CM110/112:
<Status Message>
<Query Byte> Status Message
<00>
D
<Position High Byte> <Low Byte>
<01>
D
<0> <Type Byte>
0 = Single
1 = Additive dbl(CM112 only)
254 = Subtractive dbl(CM112 only)
<02>
D
<Grooves/mm High Byte> <Low Byte>
<03>
D
<Blaze High Byte> <Low Byte>
<04>
D
<0> <Current Grating No.>
<05>
D
<Speed High Byte> <Low Byte>
<06>
D
<0> <Size Byte>
<13>
D
<0> <Number of gratings>
<14>
D
<0>
<Current
Units>
<19>
D
<Serial Number High Byte> <Serial Number Low Byte>
CM110/112 Action: No action.
From CM110/112:
<Status Byte>
From CM110/112:
<24>
D
RESET
This command returns the grating to home position.
To CM110/112:
<255>
D
<255><255>
CM110/112 Action: Grating will return to home position
SCAN
This command scans the monochromator between a START position and an END position at a rate determined by the
Spectral ProductsEED command. The START may be greater or smaller than the END. Valid values of position are
grating and units dependent and are described in Appendix B, page 21.
To CM110/112:
<12>
D
<Start High Byte> <Start Low Byte>
<End High Byte> <End Low Byte>
From CM110/112:
<Status Byte>
CM110/112 Action: Sets CTS low, and moves below (above) the starting value. Sets CTS high and accelerates to
scanning speed. Sets CTS low when the Start is reached. Sets CTS high again as END is passed.
Spectral Products
2659-A Pan American Frwy., NE Albuquerque, NM 87107 Tel (505) 343-9700 Fax (505) 343-9705
www.spectralproducts.com
9
From CM110/112:
<24>
D
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com